A Multicenter Evaluation of a Closed-Loop Anesthesia Delivery System: A Randomized Controlled Trial.

نویسندگان

  • Goverdhan D Puri
  • Preethy J Mathew
  • Indranil Biswas
  • Amitabh Dutta
  • Jayashree Sood
  • Satinder Gombar
  • Sanjeev Palta
  • Morup Tsering
  • P L Gautam
  • Aveek Jayant
  • Inderjeet Arora
  • Vishal Bajaj
  • T S Punia
  • Gurjit Singh
چکیده

BACKGROUND Closed-loop systems for anesthesia delivery have been shown to outperform traditional manual control in different clinical settings. The present trial was aimed at evaluating the feasibility and efficacy of Bispectral Index (BIS)-guided closed-loop anesthesia delivery system (CLADS) in comparison with manual control across multiple centers in India. METHODS Adult patients scheduled for major surgical procedures of an expected duration of 1 to 3 hours were randomized across 6 sites into 2 groups: a CLADS group and a manual group. In the manual control group, propofol infusion was titrated manually by the attending anesthesiologist to a BIS of 50 during induction and maintenance. Analgesia was maintained with fentanyl infusion and nitrous oxide in both groups. In the CLADS group, both induction and maintenance of anesthesia were performed automatically using CLADS. The primary outcome measure was the performance of the system as assessed by the percentage of total anesthesia time BIS remained ±10 of target BIS. The secondary outcome measures were a percentage of anesthesia-time heart rate and mean arterial pressure within 25% of the baseline, median absolute performance error, wobble, and global score. Wobble indicates intraindividual variability in the control of BIS, and global score reflects the overall performance; lower values indicate superior performance for both parameters. The performance parameters of the system also were compared among the participating sites. RESULTS Two hundred forty-two patients were randomized. BIS was maintained within ±10 of target for significantly longer time in the CLADS group (81.4% ± 8.9 % of anesthesia duration) than in the manual group (55.34% ± 25%, P < 0.0001). The indices that assess performance were significantly better in the CLADS group than the manual group as follows: median absolute performance error was 10 (10, 12) (median [interquartile range]) in the CLADS group versus 18 (14, 24) in the manual group, P < 0.0001; wobble was 9 (8, 10) in CLADS group versus 10 (8, 14) in the manual group, P = 0.0009; and Global score, which reflects overall performance, was 24 (19, 30) in the CLADS group versus 51 (31, 99) in the manual group, P < 0.0001. The percentage of time heart rate was within 25% of the baseline was significantly greater in the CLADS group (heart rate of 95 [87, 99], median [interquartile range], in the CLADS group versus 90 [75, 98] in the manual group P = 0.0031). On comparison of data between the centers, the performance parameters did not differ significantly among the centers in the CLADS group (P = 0.94), but the parameters differed significantly among the centers in the manual group (P < 0.001). CONCLUSIONS Our study in a multicenter setting proves the consistently better performance of automated anesthesia drug delivery compared with conventional manual control. This highlights an important advantage of an automated system for delivering standardized anesthesia, thereby overcoming differences in practices among anesthesiologists.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System

BACKGROUND The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to compare closed-loop control and open-loop control of propofol by using the CONCERT-CL closed-loop in...

متن کامل

QUALITY AND PATIENT SAFETY Evaluation of a novel closed-loop total intravenous anaesthesia drug delivery system: a randomized controlled trial

Editor’s key points † Automatic feedback control depends on having a reliable, measurable control variable. † An automatic system for closed-loop administration of i.v. anaesthesia drugs was compared with manual control. † Bispectral index and a compound analgesia score were used as the target controls. † Time within excellent (≤10%) and good (11–20%) range of the two targets was better with th...

متن کامل

The effect of low dose fentanyl as a premedication before induc-tion of general anesthesia on the neonatal apgar score in cesarean section delivery: randomized, double-blind controlled trial

Background: The administration of opioids before induction of general anesthesia can be considered as a problem in cesarean section. The aim of this study was to compare the effects of intravenous Fentanyl as a premedication before induction of general anesthesia versus placebo on maternal hemodynamic parameters and on the first and fifth minutes Apgar score in the neonates in elective cesarean...

متن کامل

A New Approach in Preliminary Design of Closed Loop Solar Thermal Systems (RESEARCH NOTE)

In this paper, a model for closed loop solar system is presented and an attempt is made to generalize the model to be utilized for primary design of any solar active thermal system. This model may be used for systems in which gas or a liquid are fluids that flow. Two new parameters, namely, the system heat delivery factor and the system heat absorption factor are introduced in the model. These ...

متن کامل

Comparing the Prophylactic Effect of Pregabalin with that of Aminophylline Plus Dexamethasone on Postdural Puncture Headache after Spinal Anesthesia in Cesarean Section: A Double-Blind Controlled Randomized Clinical Trial

Background: Headache is yet considered an undesirable complication of spinal anesthesia as a selective method in cesarean section. The present study was conducted to compare the prophylactic effect of pregabalin with that of aminophylline plus dexamethasone on postdural puncture headache after spinal anesthesia in cesarean section. Materials and Methods: The present double-blind controlled rand...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesia and analgesia

دوره 122 1  شماره 

صفحات  -

تاریخ انتشار 2016