Validation of a Fiducial Frame System for Image-guided Otologic Surgery Utilizing Baha® Bone Screws
نویسندگان
چکیده
Image-guided otologic surgeries performed on healthy patients require non-invasive fiducial systems that achieve submillimetric target registration error (TRE). We have described such a system that attaches repeatably to a subject via a dental bite block [1]. We now describe a new method for the clinical validation of such a system that can be conducted in the laboratory with no intraoperative data collection. The method relies on the participation of patients with bone-anchored hearing-aids (BAHA®), which require the implantation of a titanium screw behind the ear. By attaching a fiducial marker to this screw, we can measure TRE at the screw with high accuracy. From those measurements we can estimate TRE at other positions. Results of measurements on a cadaveric skull agree with a previous study that required the implantation of intracranial posts [9] and show that the system achieves submillimetric accuracy in the inner ear.
منابع مشابه
Clinical Determination of Target Registration Error of an Image- Guided Otologic Surgical System using Patients with Bone-Anchored Hearing Aids
Image guidance in otologic surgery has been thwarted by the need for a non-invasive fiducial system with target registration error (TRE) at the inner ear below 1.5mm. We previously presented a fiducial frame for this purpose that attaches to the upper dentition via patient-specific bite blocks and demonstrated a TRE of 0.73mm (±0.23mm) on cadaveric skulls. In that study, TRE measurement depende...
متن کاملSubmillimetric target-registration error using a novel, non-invasive fiducial system for image-guided otologic surgery.
OBJECTIVE Otologic surgery is undertaken to treat ailments of the ear, including persistent infections, hearing loss, vertigo, and cancer. Typically performed on otherwise-healthy patients in outpatient facilities, the application of image-guided surgery (IGS) has been limited because accurate (<1 mm), non-invasive fiducial systems for otologic surgery have not been available. We now present su...
متن کاملBite-Block Relocation Error in Image-Guided Otologic Surgery
Otologic surgery is undertaken to treat ailments of the ear including persistent infections, hearing loss, vertigo, and cancer. Typically performed on healthy patients in outpatient facilities, the application of image-guided surgery has been limited because accurate (<1mm), non-invasive fiducial systems for otologic surgery are not available. We have developed such a system, which repeatably a...
متن کاملMinimally invasive, image-guided, facial-recess approach to the middle ear: demonstration of the concept of percutaneous cochlear access in vitro.
HYPOTHESIS Image-guided surgery will permit accurate access to the middle ear via the facial recess using a single drill hole from the lateral aspect of the mastoid cortex. BACKGROUND The widespread use of image-guided methods in otologic surgery has been limited by the need for a system that achieves the necessary level of accuracy with an easy-to-use, noninvasive fiducial marker system. We ...
متن کاملImage-guided otologic surgery
Application of image-guided surgical systems to otologic surgery has been limited by the need for submillimeter accuracy via a fiducial system that is easily useable (i.e. noninvasive and nonobstructive). To overcome these limitations and demonstrate the feasibility of image-guided otologic surgery, a novel fiducial marker system was constructed which attaches to a dental biteblock and has fidu...
متن کامل