Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer
نویسندگان
چکیده
INTRODUCTION Hemoglobin concentration is assessed to detect anemia and its associated morbidities. Hemoglobin is usually determined from venous or capillary blood samples run on a laboratory analyzer. However, this method requires a needle stick and results can be delayed. It also exposes caregivers to risks associated with needle sticks and blood exposure. Noninvasive hemoglobin determination would be of benefit to patients and caregivers because it would allow for quick and painless point-of-care assessment. METHODS Hemoglobin determination from a noninvasive spot check hemoglobin device (Pronto-7 with SpHb, Masimo) and an invasive point-of-care device (HemoCue) was compared with venous blood samples run on a laboratory hematology analyzer. RESULTS A total of 440 outpatients and healthy volunteers were included (mean age 36 years, 62% female). Compared with the hematology analyzer, the bias ± standard deviation of was -0.1 ± 1.1 g/dL for SpHb and -0.1 ± 1.6 g/dL for HemoCue. CONCLUSION Noninvasive hemoglobin testing with SpHb provided similar accuracy as invasive point-of-care hemoglobin testing and may enable more efficient and effective patient care.
منابع مشابه
Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients.
OBJECTIVE To determine whether noninvasive hemoglobin measurement by Pulse CO-Oximetry could provide clinically acceptable absolute and trend accuracy in critically ill patients, compared to other invasive methods of hemoglobin assessment available at bedside and the gold standard, the laboratory analyzer. DESIGN Prospective study. SETTING Surgical intensive care unit of a university teachi...
متن کاملAssessing serum hemoglobin levels without venipuncture: accuracy and reliability of Pronto-7 noninvasive spot-check device.
BACKGROUND Hemoglobin is a frequently obtained test in hospital settings. We analyzed accuracy of a noninvasive device compared to standard laboratory analyzers in a variety of settings. METHODS A noninvasive hemoglobin monitoring device was analyzed for reliability, correlation, precision, and bias. Hemoglobin levels were obtained from standard laboratory and point-of-care hemoglobin analyze...
متن کاملThe efficacy of noninvasive hemoglobin measurement by pulse CO-oximetry in neonates.
OBJECTIVE To evaluate clinical applicability of noninvasive hemoglobin (Hb) measurement with a pulse CO-oximeter in neonates. DESIGN Prospective comparison study. SETTING Neonatal ICU. PATIENTS Fifty-six preterm and term infants with median age = 20 days (range = 1-98 days) and median weight = 1,440 g (range = 530-4,230 g). INTERVENTIONS Hb measurements by Pulse CO-Oximetry (Masimo Radi...
متن کاملEvaluation of the Afinion AS100 Point-of-Care Analyzer for Hemoglobin A1c
An evaluation of a new Afinion AS100 Analyzer was conducted to assess analytical performance. Precision was estimated by analyzing 2 control and 3 patient samples twice a day for 10 days. Accuracy was established by analysis of 6 samples from the National Glycated Hemoglobin Standardization Program for a 3-day period. Agreement was correlated to a laboratory method, the Variant II Turbo Hemoglo...
متن کاملConsiderations for evaluating the accuracy of hemoglobin monitoring.
To the Editor: Masimo manufactures the Radical-7, a multi-wavelength Pulse CO-Oximeter that continuously measures noninvasive hemoglobin concentration (SpHb). This technology was the subject of a study by Applegate et al., evaluating the accuracy of revision “E” SpHb sensors and software during abdominal and pelvic surgery. Masimo appreciates the work of Applegate et al., and we are grateful fo...