Manual versus Mechanical Chest Compressions on Surfaces of Varying Softness with or without Backboards: A Randomized, Crossover Manikin Study.

نویسندگان

  • Gabriel Putzer
  • Anna Fiala
  • Patrick Braun
  • Sabrina Neururer
  • Karin Biechl
  • Bernhard Keilig
  • Werner Ploner
  • Ernst Fop
  • Peter Paal
چکیده

BACKGROUND Chest compression quality is decisive for overall outcome after cardiac arrest. Chest compression depth may decrease when cardiopulmonary resuscitation (CPR) is performed on a mattress, and the use of a backboard does not necessarily improve compression depth. Mechanical chest compression devices may overcome this problem. OBJECTIVES We sought to investigate the effectiveness of manual chest compressions both with and without a backboard compared to mechanical CPR performed on surfaces of different softness. METHODS Twenty-four advanced life support (ALS)-certified rescuers were enrolled. LUCAS2 (Physio-Control, Redmond, WA) delivers 52 ± 2 mm deep chest compressions and active decompressions back to the neutral position (frequency 102 min(-1); duty cycle, 50%). This simulated CPR scenario was performed on a Resusci-Anne manikin (Laerdal, Stavanger, Norway) that was lying on 3 different surfaces: 1) a concrete floor, 2) a firm standard mattress, and 3) a pressure-relieving mattress. Data were recorded by the Laerdal Skill Reporting System. RESULTS Manual chest compression with or without a backboard were performed correctly less often than mechanical chest compressions (floor: 33% [interquartile range {IQR}, 27-48%] vs. 90% [IQR, 86-94%], p < 0.001; standard mattress: 32% [IQR, 20-45%] vs. 27% [IQR, 14-46%] vs. 91% [IQR, 51-94%], p < 0.001; and pressure-relieving mattress 29% [IQR, 17-49%] vs. 30% [IQR, 17-52%] vs. 91% [IQR, 87-95%], p < 0.001). The mean compression depth on both mattresses was deeper with mechanical chest compressions (floor: 53 mm [range, 47-57 mm] vs. 56 mm [range, 54-57 mm], p = 0.003; standard mattress: 50 mm [range, 44-55 mm] vs. 51 mm [range, 47-55 mm] vs. 55 mm [range, 54-58 mm], p < 0.001; and pressure-relieving mattress: 49 mm [range, 44-55 mm] vs. 50 mm [range, 44-53 mm] vs. 55 mm [range, 55-56 mm], p < 0.001). In this ∼6-min scenario, the mean hands-off time was ∼15 to 20 s shorter in the manual CPR scenarios. CONCLUSIONS In this experimental study, only ∼30% of manual chest compressions were performed correctly compared to ∼90% of mechanical chest compressions, regardless of the underlying surface. Backboard use did not influence the mean compression depth during manual CPR. Chest compressions were deeper with mechanical CPR. The mean hands-off time was shorter with manual CPR.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Mechanical versus manual chest compressions for out-of-hospital cardiac arrest: a meta-analysis of randomized controlled trials

Recent evidence regarding mechanical chest compressions in out-of-hospital cardiac arrest (OHCA) is conflicting. The objective of this study was to perform a meta-analysis of randomized controlled trials (RCTs) to compare the effect of mechanical versus manual chest compressions on resuscitation outcomes in OHCA. PubMed, Embase, the Cochrane Central Register of Controlled Trials, and the Clinic...

متن کامل

Comparison of Pentax-AWS Airwayscope and Glidescope for Infant Tracheal Intubation by Anesthesiologists during Cardiopulmonary Arrest Simulation: A Randomized Crossover Trial

BACKGROUND Recent guidelines for infant cardiopulmonary resuscitation emphasize that all rescuers should minimize interruption of chest compressions, even for endotracheal intubation. We compared the utility of the Pentax-AWS Airwayscope (AWS) with the Glidescope (GS) during chest compressions on an infant manikin. METHODS Twenty-four anesthesiologists with more than two years of experience p...

متن کامل

Can the ETView VivaSight SL Rival Conventional Intubation Using the Macintosh Laryngoscope During Adult Resuscitation by Novice Physicians?

UNLABELLED The aim of this study was to assess the performance of the ETView VivaSight SL (ETView) single-lumen airway tube with an integrated high-resolution imaging camera in a manikin-simulated cardiopulmonary resuscitation scenario with and without chest compression. This was a randomized crossover manikin trial. Following a brief training session, 107 volunteer novice physicians who were i...

متن کامل

Use of a Metronome in Cardiopulmonary Resuscitation: A Simulation Study.

OBJECTIVE Determine whether the use of a metronome improves chest compression rate and depth during cardiopulmonary resuscitation (CPR) on a pediatric manikin. METHODS A prospective, simulation-based, crossover, randomized controlled trial was conducted. Participants included pediatric residents, fellows, nurses, and medical students who were randomly assigned to perform chest compressions on...

متن کامل

Comparison of the quality of chest compressions on a dressed versus an undressed manikin: A controlled, randomised, cross-over simulation study

BACKGROUND Undressing the chest of a cardiac arrest victim may delay the initiation of chest compressions. Furthermore, expecting laypeople to undress the chest may increase bystander reluctance to perform cardiopulmonary resuscitation (CPR). Both of these factors might conceivably decrease survival following cardiac arrest. Therefore, the aim of this study was to examine if the presence or abs...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of emergency medicine

دوره 50 4  شماره 

صفحات  -

تاریخ انتشار 2016