Tubing failure during prolonged roller pump use: a laboratory study.

نویسندگان

  • G J Peek
  • K Wong
  • C Morrison
  • H M Killer
  • R K Firmin
چکیده

Little is known about the mechanical forces acting on extracorporeal circuit tubing with prolonged roller pump use during extracorporeal membrane oxygenation (ECMO). We examined the time to tubing rupture of three different materials during actual roller pump use, mean and standard deviation (SD) (SD shown in parentheses): Tygon (control) 243.7 h (175.4); LVA 121 h (14.3); and SRT 6.6 h (2.1). Failure times for both LVA and SRT were significantly different from the control (paired t-test, p = 0.02 and p < 0.001, respectively). The minimum failure times for Tygon and LVA were 99 and 101 h, respectively. We then examined Tygon under conditions of pure compression, demonstrating that even after 3.67 million compression cycles at full occlusion crack formation did not occur. If the tubing was over-occluded, cracks appeared within 24 h. Scanning electron microscopy of Tygon, which has been used during clinical ECMO, and the failure pattern during destruction testing demonstrate that shear stress and compression coexist during clinical ECMO. Use of under-occlusive pump settings could improve tubing life.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Tygon" Roller Pump Tubing

A failure analysis was performed on Norton S-65-HL Tygon® tubing. Fatigue testing was performed on four sizes of this tubing, and essentially showed how the tubing wears out. A dynamic "life hours to failure" test, which was performed on the 3/8" internal diameter (ID) size, quantified when the tubing ruptured. Based on results of laboratory testing and the institution's clinical extracorporeal...

متن کامل

Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass.

Roller pumps are widely used for cardiopulmonary bypass in developing nations by virtue of proven safety during several years of institutional use and cost effectiveness. However, careful adjustment of roller occlusion is needed because they are known to cause spallation, tubing wear, and the occasional incident of rupture of tubing in the extracorporeal circuit. Rupture of polyvinylchloride tu...

متن کامل

Roller Pump Induced Tubing Wear: Another Argument in Favor of Arterial Line Filtration

____________ _ Pump head tubing was studied to determine if internal wear occurred during routine cardiopulmonary bypass. Several segments of silicone rubber tubing were examined in a scanning electron microscope following clinical perfusion. In each case, the proper degree of occlusion had been set just prior to beginning bypass. The roller pump was a six-inch dual roller type, and the tubing ...

متن کامل

Can vacuum assisted venous drainage be achieved using a roller pump in an emergency? A pilot study using neonatal circuitry.

There has been much advancement in perfusion technology over its 50 years of progression. One of these techniques is vacuum-assisted venous drainage (VAVD). Many perfusionists augment venous drainage using VAVD, typically from a wall vacuum source. This study explores alternates to providing VAVD if the wall vacuum fails. In two porcine laboratories, approximately 36 in. of 3/16-in. tubing was ...

متن کامل

The use of plastic tubing in roller pumps.

If plastic tubing is to be used in roller pumps for propelling blood in extracorporeal circulations under occlusive conditions spring loading the rollers and accurately setting the spring force is essential. Various types and sizes of plastic tubing are compared and their performance estimated; rubber tubing used for the same purpose has been tested. The optimum spring force has been determined...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Perfusion

دوره 14 6  شماره 

صفحات  -

تاریخ انتشار 1999