The quality of laboratory testing today: an assessment of sigma metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance.

نویسندگان

  • James O Westgard
  • Sten A Westgard
چکیده

To assess the analytic quality of laboratory testing in the United States, we obtained proficiency testing survey results from several national programs that comply with Clinical Laboratory Improvement Amendments (CLIA) regulations. We studied regulated tests (cholesterol, glucose, calcium, fibrinogen, and prothrombin time) and nonregulated tests (international normalized ratio [INR], glycohemoglobin, and prostate-specific antigen [PSA]). Quality was assessed on the sigma scale with a benchmark for minimum process performance of 3 sigma and a goal for world-class quality of 6 sigma. Based on the CLIA criteria for acceptable performance in proficiency testing (allowable total errors [TEa]), the national quality of cholesterol testing (TEa = 10%) estimated sigma values as 2.9 to 3.0; glucose (TEa = 10%), 2.9 to 3.3; calcium (TEa = 1.0 mg/dL), 2.8 to 3.0; prothrombin time (TEa = 15%), 1.8; INR (TEa = 20%), 2.4 to 3.5; fibrinogen (TEa = 20%), 1.8 to 3.2; glycohemoglobin (TEa = 10%), 1.9 to 2.6; and PSA (TEa = 10%), 1.2 to 1.8. The analytic quality of laboratory tests requires improvement in measurement performance and more intensive quality control monitoring than the CLIA minimum of 2 levels per day.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA Criteria for Acceptable Performance

To assess the analytic quality of laboratory testing in the United States, we obtained proficiency testing survey results from several national programs that comply with Clinical Laboratory Improvement Amendments (CLIA) regulations. We studied regulated tests (cholesterol, glucose, calcium, fibrinogen, and prothrombin time) and nonregulated tests (international normalized ratio [INR], glycohemo...

متن کامل

Quantitative Assessment of Analytical Phase Quality of Clinical Biochemistry Parameters Using Sigma Metrics

Background and Aims: Six sigma is the latest version of total quality management. It is quantitative goal for process performance. With increasing demands for improved accuracy and reliability of the results, Six Sigma is gaining increased visibility in the clinical laboratory process outcomes. The aim of study was to evaluate the quality of analytical phase performance in a clinical biochemist...

متن کامل

How good are clinical laboratories? An assessment of current performance.

The Clinical Laboratory Improvement Act of 1967 and Amendments of 1988 (CLIA '67 and CLIA '88) were enacted to ensure that clinical laboratories within the U.S. provide a quality of service that meets clinical needs for good patient care. Approved proficiency-testing programs are to judge the quality of laboratory testing by promulgated performance criteria. We examine the quality of analytical...

متن کامل

Evaluation of biochemical methods performance in clinical laboratories: is this performance suitable for clinical application?

Background: Clinical application of the test results are profoundly influenced by laboratory errors. Nowadays, the best way to evaluate the performance of a method is to determine its sigma scale according to total allowable error (TEa) and analytical errors, including random and systematic errors. In this study, the performance of the measuring kits for biochemical analytes belonging to Pars A...

متن کامل

Ranking Iranian Accredited Laboratories with Proficiency Tests Using PR0METHEE Method

The rapid pace of changing technology is leading to high rate of import and export between different countries. In outsourcing governmental organizations duties to private sectors, it is a need to control sectors and they should be under assessment policy. This research has been developed to show how the accredited laboratories of Iranian standard (ISIRI) can be assessed for their technical ser...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • American journal of clinical pathology

دوره 125 3  شماره 

صفحات  -

تاریخ انتشار 2006