MRI-compatible voltage-based electroanatomic mapping system for 3T MR-guided cardiac electrophysiology: swine validations
نویسندگان
چکیده
Background MRI provides luminal, edema, & scar images which assist in the Electrophysiological (EP) treatment of ventricular and atrial arrhythmias]. Until a complete set of MR-compatible EP-devices is available, patients must be repeatedly moved between the MRI, where imaging and mapping occur, to the conventional EP suite, where puncture, navigation and Radio-Frequency Ablation occur. MRI-conditional voltage-based electroanatomic mapping (EAM) would permit efficient MRI-guided EP, with registration-free continuation outside MRI, utilizing X-ray, Intra-Cardiac-Echo (ICE) and EAM guidance. A 1.5T EnSiteTM VelocityTM cardiac mapping system, a voltage-based EAM available from St. Jude Medical was previously validated [1,2]. Multiple-catheter EAM (localization and intra-cardiac electrocardiogram (EGM) measurement) inside a 3T MRI requires modifications.
منابع مشابه
Voltage-based electroanatomic mapping system for MR-guided cardiac electrophysiology: preliminary swine validations
Background MRI produces images that serve as luminal, edema, & scar maps to assist in the Electrophysiological (EP) treatment of ventricular and atrial arrhythmias [1]. Until MR-compatible EP devices are widely available, there will be a need to perform EP partially in the MRI for imaging, and partially outside the MRI for ablation, puncture & navigation. An MR-conditional voltage-based Electro...
متن کاملElectroanatomic mapping and radiofrequency ablation of porcine left atria and atrioventricular nodes using magnetic resonance catheter tracking.
BACKGROUND The MRI-compatible electrophysiology system previously used for MR-guided left ventricular electroanatomic mapping was enhanced with improved MR tracking, an MR-compatible radiofrequency ablation system and higher-resolution imaging sequences to enable mapping, ablation, and ablation monitoring in smaller cardiac structures. MR-tracked navigation was performed to the left atrium (LA)...
متن کاملVURTIGO: Visualization Platform for Real-Time, MRI-Guided Cardiac Electroanatomic Mapping
Guidance of electrophysiological (EP) procedures by magnetic resonance imaging (MRI) has significant advantages over x-ray fluoroscopy. Display of electroanatomic mapping (EAM) during an intervention fused with a prior MR volume and DE-MRI derived tissue classification should improve the accuracy of cardiac resynchronization therapy (CRT) for ventricular arrhythmias. Improved accuracy in the sp...
متن کاملVoltage-based device tracking in a 1.5 Tesla MRI during imaging: initial validation in swine models.
PURPOSE Voltage-based device-tracking (VDT) systems are commonly used for tracking invasive devices in electrophysiological cardiac-arrhythmia therapy. During electrophysiological procedures, electro-anatomic mapping workstations provide guidance by integrating VDT location and intracardiac electrocardiogram information with X-ray, computerized tomography, ultrasound, and MR images. MR assists ...
متن کاملElectroanatomic mapping of the left ventricle in a porcine model of chronic myocardial infarction with magnetic resonance-based catheter tracking.
BACKGROUND X-ray fluoroscopy constitutes the fundamental imaging modality for catheter visualization during interventional electrophysiology procedures. The minimal tissue discriminative capability of fluoroscopy is mitigated in part by the use of electroanatomic mapping systems and enhanced by the integration of preacquired 3-dimensional imaging of the heart with computed tomographic or magnet...