Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners.

نویسندگان

  • A E Bowden
  • S M Kurtz
  • A A Edidin
چکیده

In this study, a novel micro-CT-based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high-resolution micro-CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16-bit precision) with the use of a 1,024 x 1,024 in-plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm(3)/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra-- and interobserver uncertainty analysis showed the method to be both accurate and repeatable.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Highly cross-linked polyethylene acetabular liners retrieved four to five years after revision surgery: a report of two cases.

There is currently considerable interest in the use of highly cross-linked polyethylene (XLPE) acetabular liners for total hip arthroplasty (THA). In literature, only a single retrieval analysis of one type of XLPE liner implanted for greater than four years exists. The purpose of the present report is to quantify surface deviations in two XLPE liners implanted during revision THA and retrieved...

متن کامل

Backside volumetric change in the polyethylene of uncemented acetabular components.

Polyethylene wear of acetabular components is a key factor in the development of periprosthetic osteolysis and wear at the articular surface has been well documented and quantified, but fewer data are available about changes which occur at the backside of the liner. At revision surgery for loosening of the femoral component we retrieved 35 conventional modular acetabular liners of the same desi...

متن کامل

In vivo and ex vivo measurement of polyethylene wear in total hip arthroplasty

BACKGROUND Determination of the amount of wear in a polyethylene liner following total hip arthroplasty (THA) is important for both the clinical care of individual patients and the development of new types of liners. PATIENTS AND METHODS We measured in vivo wear of the polyethylene liner using computed tomography (CT) (obtained in the course of regular clinical care) and compared it to coordi...

متن کامل

Retrieval analysis of Harris-Galante I and II acetabular liners in situ for more than 10 years

BACKGROUND AND PURPOSE There have been few reports documenting the wear and oxidation performance of the polyethylene bearing surface of HGPI and HGPII THA devices. We evaluated retrieved HGPI and HGPII acetabular liners that had been in situ for more than 10 years and determined whether there was a relationship between clinical and radiographic factors, surface damage, wear, and oxidation. M...

متن کامل

Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro

Backside wear due to micromotion and poor conformity between the liner and its titanium alloy shell may contribute to the high rates of retroacetabular osteolysis and consequent aseptic loosening. The purpose of our study was to understand the wear process on the backside of polyethylene liners from two acetabular cup systems, whose locking mechanism is based on a press-fit cone in combination ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of biomedical materials research. Part B, Applied biomaterials

دوره 75 1  شماره 

صفحات  -

تاریخ انتشار 2005