Shaping ability of Reciproc and TF Adaptive systems in severely curved canals of rapid microCT-based prototyping molar replicas
نویسندگان
چکیده
OBJECTIVE To evaluate the shaping ability of Reciproc and Twisted-File Adaptive systems in rapid prototyping replicas. MATERIAL AND METHODS Two mandibular molars showing S-shaped and 62-degree curvatures in the mesial root were scanned by using a microcomputed tomography (μCT) system. The data were exported in the stereolitograhic format and 20 samples of each molar were printed at 16 µm resolution. The mesial canals of 10 replicas of each specimen were prepared with each system. Transportation was measured by overlapping radiographs taken before and after preparation and resin thickness after instrumentation was measured by μCT. RESULTS Both systems maintained the original shape of the apical third in both anatomies (P>0.05). Overall, considering the resin thickness in the 62-degree replicas, no statistical difference was found between the systems (P>0.05). In the S-shaped curvature replica, Reciproc significantly decreased the thickness of the resin walls in comparison with TF Adaptive. CONCLUSIONS The evaluated systems were able to maintain the original shape at the apical third of severely curved mesial canals of molar replicas.
منابع مشابه
Comparison of Shaping Ability of Five Nickel-Titanium Rotary Instruments in Simulated Curved Canals
This study compared the shaping ability of five Ni-Ti rotary files (ProTaper Universal, Reciproc, Hyflex CM, ProTaper Next and TF Adaptive) in simulated root canals. Each group consisted of 10 root canals. Preand post-instrumentation photographs were taken using a precise camera and superimposed with Photoshop. Material removal from the inner and outer canal walls, transportation and centering ...
متن کاملمقایسه ترانسپورتیشن دو سیستم روتاری رسیپروک و نئولیکس در کانالهای انحنادار ریشه دندان توسط توموگرافی کامپیوتری با اشعه مخروطی
Background and Objective: The preparation of the root canal (i.e., cleaning and shaping) is the main step in the endodontic treatment. The nickel-titanium rotary files result in fewer errors during the preparation of curved canals. This study aimed to compare the transportation of Neolix and Reciproc rotary systems in curved canals using cone beam computed tomography (CBCT). Materials and Meth...
متن کاملShaping ability of reciprocating single-file and full-sequence rotary instrumentation systems in simulated curved canals
OBJECTIVE The purpose of this study was to evaluate the shaping ability of three nickel-titanium systems in simulated curved canals. MATERIALS AND METHODS Sixty simulated canals were prepared to apical size 25 with Reciproc, S5, and twisted file (TF) instruments. Standardized pre and postoperative images were taken using a digital camera, were superimposed and aberrations were recorded. Mater...
متن کاملShaping Ability of NiTi Rotary versus Stainless-Steel Hand Instruments in Curved Root Canals of Extracted Teeth
Introduction: The purpose of this in-vitro study was to compare shaping ability of NiTi rotary Mtwo and FlexMaster with stainless steel hand K-Flexofile in curved root canals of extracted teeth. Materials and Methods: Mesiobuccal canals of 45 mandibular molar teeth were balanced into three groups of 15 each, with respect to the Schneider angle, canal access angle, height, dis-tance and radiu...
متن کاملReciproc : Ein neuer Ansatz zu einem alten Problem
[1] Ahmad M. Comparison between two ultrasonic units in shaping curved canals in teeth using macroradiography. cleaning effectiveness of two single-file systems in severely curved root canals of extracted teeth: Reciproc WaveOne versus Mtwo ProTaper. Fidel RAS. The ability of the Reciproc R25 instrument to reach the full root canal working length without a glide path.
متن کامل