Accuracy of dual‐energy computed tomography for the quantification of iodine in a soft tissue‐mimicking phantom

نویسندگان

  • Jung‐Hui Li
  • Yeh‐Ming Du
  • Hsuan‐Ming Huang
چکیده

The objective of this study was to evaluate the accuracy of dual-energy CT (DECT) for quantifying iodine using a soft tissue-mimicking phantom across various DECT acquisition parameters and dual-source CT (DSCT) scanners. A phantom was constructed with plastic tubes containing soft tissue-mimicking materials with known iodine concentrations (0-20 mg/mL). Experiments were performed on two DSCT scanners, one equipped with an integrated detector and the other with a conventional detector. DECT data were acquired using two DE modes (80 kV/Sn140 kV and 100 kV/Sn140 kV) with four pitch values (0.6, 0.8, 1.0, and 1.2). Images were reconstructed using a soft tissue kernel with and without beam hardening correction (BHC) for iodine. Using the dedicated DE software, iodine concentrations were measured and compared to true concentrations. We also investigated the effect of reducing gantry rotation time on the DECT-based iodine measurement. At iodine concentrations higher than 10 mg/mL, the relative error in measured iodine concentration increased slightly. This error can be decreased by using the kernel with BHC, compared with the kernel without BHC. Both 80 kV/Sn140 kV and 100 kV/Sn140 kV modes could provide accurate quantification of iodine content. Increasing pitch value or reducing gantry rotation time had only a minor impact on the DECT-based iodine measurement. The DSCT scanner, equipped with the new integrated detector, showed more accurate iodine quantification for all iodine concentrations higher than 10 mg/mL. An accurate quantification of iodine can be obtained using the second-generation DSCT scanner in various DE modes with pitch values up to 1.2 and gantry rotation time down to 0.28 s. For iodine concentrations ≥ 10 mg/mL, using the new integrated detector and the kernel with BHC can improve the accuracy of DECT-based iodine measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values for four computed tomography scanners

Objective: In this study, we aimed to validate the accuracy of computed tomography-based attenuation correction (CTAC) using the bilinear scaling method.Methods: The measured attenuation coefficient (μm) was compared to a theoretical attenuation coefficient (μt ) using four different CT scanners and an RMI 467 phantom. The effective energy of the CT beam X-rays was calculated, using the aluminu...

متن کامل

Analysis of the Attenuation Coefficient of Composite Silicone Rubber and Gliceryn for Soft Tissue Phantom Applications

Introduction: Phantom is an object that can be used to investigate the accuracy of radiation dose delivered to patients. Phantom is usually produced from the combination of the silicone rubber as a matrix and glycerin as a filler to form a composite for the replacement of the human soft tissue. The composite is imaged using computed tomography (CT) simulator for the de...

متن کامل

Development of a Liver Phantom Based on Computed Tomography Images for Dosimetric Purpose

Introduction: The present study was conducted with the aim of designing a liver phantom for dosimetry. To benchmark the results obtained by the developed liver phantom, another method was applied for the dosimetry of a real liver tissue using imaging. Materials and Methods: For the purpose of the study, a real liver tissue was converted into a phantom based on thegram-molecular weight of the co...

متن کامل

Disseminated thoracoabdominal splenosis mimicking metastatic disease: A case of colorectal cancer

Here, we describe a patient with a history of colorectal cancer in whom 2-[18F]fluoro-2-deoxy-D-glucose positron emission tomography/computed tomography ([18F]FDG PET/CT) was performed for the evaluation of response to therapy. [18F]FDG PET/CT showed a small residual disease in the rectum. In addition, multiple metabolically inactive soft tissue densities were d...

متن کامل

Correlation Between Tissue Densities in Computed Tomography and Three Different Cone-Beam Computed Tomography Units (In Vitro)

Background and Aim: Bone density is of great assistance in the selection of the proper implant site. The present study aimed to assess the correlation between tissue densities in computed tomography (CT) and three different cone-beam computed tomography (CBCT) units. Materials and Methods: In this descriptive study, a radiographic phantom consisting of a transparent polymethyl methacrylate (PM...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015