Registration error of the liver CT using deformable image registration of MIM Maestro and Velocity AI

نویسندگان

  • Nobuyoshi Fukumitsu
  • Kazunori Nitta
  • Toshiyuki Terunuma
  • Toshiyuki Okumura
  • Haruko Numajiri
  • Yoshiko Oshiro
  • Kayoko Ohnishi
  • Masashi Mizumoto
  • Teruhito Aihara
  • Hitoshi Ishikawa
  • Koji Tsuboi
  • Hideyuki Sakurai
چکیده

BACKGROUND Understanding the irradiated area and dose correctly is important for the reirradiation of organs that deform after irradiation, such as the liver. We investigated the spatial registration error using the deformable image registration (DIR) software products MIM Maestro (MIM) and Velocity AI (Velocity). METHODS Image registration of pretreatment computed tomography (CT) and posttreatment CT was performed in 24 patients with liver tumors. All the patients received proton beam therapy, and the follow-up period was 4-14 (median: 10) months. We performed DIR of the pretreatment CT and compared it with that of the posttreatment CT by calculating the dislocation of metallic markers (implanted close to the tumors). RESULTS The fiducial registration error was comparable in both products: 0.4-32.9 (9.3 ± 9.9) mm for MIM and 0.5-38.6 (11.0 ± 10.0) mm for Velocity, and correlated with the tumor diameter for MIM (r = 0.69, P = 0.002) and for Velocity (r = 0.68, P = 0.0003). Regarding the enhancement effect, the fiducial registration error was 1.0-24.9 (7.4 ± 7.7) mm for MIM and 0.3-29.6 (8.9 ± 7.2) mm for Velocity, which is shorter than that of plain CT (P = 0.04, for both). CONCLUSIONS The DIR performance of both MIM and Velocity is comparable with regard to the liver. The fiducial registration error of DIR depends on the tumor diameter. Furthermore, contrast-enhanced CT improves the accuracy of both MIM and Velocity. INSTITUTIONAL REVIEW BOARD APPROVAL H28-102; July 14, 2016 approved.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of deformable image registration in HDR gynecological brachytherapy

Introduction: In brachytherapy, as in external radiotherapy, image-guidance plays an important role. For GYN treatments it is standard to acquire at least CT images and preferably MR images prior to each treatment and to calculate the dose of the day on each set of images. Then, the dose to the target and to the organs at risk (OAR) is calculated with worst case scenario from I...

متن کامل

A Study on Robustness of Various Deformable Image Registration Algorithms on Image Reconstruction Using 4DCT Thoracic Images

Background: Medical image interpolation is recently introduced as a helpful tool to obtain further information via initial available images taken by tomography systems. To do this, deformable image registration algorithms are mainly utilized to perform image interpolation using tomography images.Materials and Methods: In this work, 4DCT thoracic images of five real patients provided by DI...

متن کامل

Evaluation of the tool “Reg Refine” for user‐guided deformable image registration

"Reg Refine" is a tool available in the MIM Maestro v6.4.5 platform (www.mim-software.com) that allows the user to actively participate in the deformable image registration process. The purpose of this work was to evaluate the efficacy of this tool and investigate strategies for how to apply it effectively. This was done by performing DIR on two publicly available ground-truth models, the Pixel...

متن کامل

SU-E-J-84: Is There a Threshold Limit in the Accuracy of Deformable Image Registration (DIR)? A Validation Study Using Deformable Bladder Phantom with Implanted Markers.

PURPOSE To establish the fundamental relationship between deformation and its causative physical force using a deformable bladder phantom. To ascertain if a threshold limit exists for DIR accuracy, beyond which its applicability in deformed anatomy may be clinically inappropriate. METHODS A tissue equivalent deformable bladder phantom with 21 implanted aluminum markers was developed using a v...

متن کامل

روشی جدید برای تطبیق تصاویر پزشکی هم‌نوع برپایه مدل‌های دگردیس‌پذیر و کاربرد آن برای تصاویر CT قفسه سینه

In 4D CT images of the thorax, deformation is caused by both the tissue movement and breathing. Tissue movement can be estimated at various time intervals by image registrstion. In this paper, a new deformable model is proposedfor non-regid medical image resgistration. The energy funational of the suggested algorithm consists of two internal (regulator) and external terms. By minimization of th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2017