Sensitivity of a helical diode array device to delivery errors in IMRT treatment and establishment of tolerance level for pretreatment QA

نویسندگان

  • Feliciano García‐Vicente
  • Virginia Fernández
  • Rocío Bermüdez
  • Alberto Gómez
  • Leopoldo Pérez
  • Almudena Zapatero
  • Juan J. Torres
چکیده

The aim of this study is to determine the gantry angle and multileaf collimator (MLC) gap error-detection threshold of a diode helical array with an inserted micro-ionization chamber in order to use this device for the pretreatment quality assurance (QA) of intensity-modulated radiation therapy (IMRT) treatments. Implications on the dose-volume histograms (DVHs) of the patient treatments will also be considered for the establishment of a QA protocol with a reasonable tolerance level. Three dynamic IMRT HN (head and neck) and prostate treatments were studied. Random and systematic variations of gantry angle and systematic errors in MLC gap width of the clinical treatments were analyzed in order to establish the detection sensitivity of the array. The associated clinical significance was studied introducing the same errors in the treatment plan based on the patients' computed tomography (CT) and calculating the corresponding DVHs. The Gamma (3%/3 mm) presented a 4% variation in failure rate for a rotation error of 1° for both types of treatment. Both systematic and random errors in gantry rotation angle have little effect on the patients' DVHs. MLC gap width errors of 1 mm and 2 mm in the prostate treatments imply a mean variation in isocenter-measured absorbed dose of 2.1% and 4.1%, respectively. In the case of HN, these errors entail a change in measured isocenter dose of 4.7% and 8.6%, respectively. The variation observed in the DVHs of the patients was, basically, a global displacement of the curves proportional to the isocenter dose variation caused by the gap width error. According to the array sensitivity to the analyzed errors and its implication in patient DVHs, a tolerance of 95% point passing rate for the gamma criterion 3%/2 mm and an agreement of 2% in isocenter absolute dose have been established as tolerance criteria for our pretreatment IMRT QA protocol.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

On the sensitivity of patient‐specific IMRT QA to MLC positioning errors

Accurate multileaf collimator (MLC) leaf positioning plays an essential role in the effective implementation of intensity modulated radiation therapy (IMRT). This work evaluates the sensitivity of current patient-specific IMRT quality assurance (QA) procedures to minor MLC leaf positioning errors. Random errors of up to 2 mm and systematic errors of +/-1 mm and +/-2 mm in MLC leaf positions wer...

متن کامل

Reproducibility in patient‐specific IMRT QA

The purpose of this study was to determine the reproducibility of patient-specific, intensity-modulated radiation therapy (IMRT) quality assurance (QA) results in a clinical setting. Six clinical patient plans were delivered to a variety of devices and analyses, including 1) radiographic film; 2) ion chamber; 3) 2D diode array delivered and analyzed in three different configurations (AP deliver...

متن کامل

2D linear array device as a quality assurance tool in brachytherapy applications

Background: External beam radiotherapy and brachytherapy plays a vital role in the management of cancer cervix.  High dose rate brachytherapy is being presently used worldwide for the brachytherapy applications. At present, 2-Dimensional linear array detectors   are the most common QA tool used for pretreatment patient specific quality assurance in external beam radiotherapy alon...

متن کامل

Evaluation of a 2D diode array for IMRT quality assurance.

BACKGROUND AND PURPOSE The QA of intensity modulated radiotherapy (IMRT) dosimetry is a laborious task. The goal of this work is to evaluate the dosimetric characteristics of a new 2D diode array (MapCheck from Sun Nuclear Corporation, Melbourne, Florida) and assess the role it can play in routine IMRT QA. MATERIAL AND METHODS Fundamental properties of the MapCheck such as reproducibility, li...

متن کامل

SU-E-T-123: Understanding the Meaning of IMRT QA Passing Rates with a 2D Diode Array.

PURPOSE A lower than ideal tolerance limit is used in intensity-modulated radiation therapy quality assurance (IMRT QA) with a 2D diode array due to passing rate fluctuations. The objective is to identify patterns in the passing rates to predict sources of uncertainty that can affect treatment delivery, for example, the need to re-calibrate the multileaf collimator when the passing rates start ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 13  شماره 

صفحات  -

تاریخ انتشار 2012