Physical orientation in the magnetic field affects diffusion measures: a hardware phantom study

نویسندگان

  • P. Pullens
  • A. Roebroeck
  • M. Bastiani
  • R. Goebel
چکیده

Introduction It was shown recently in an isotropic water phantom that ADC values are not equal in each measured diffusion direction due to gradient mis-calibrations [1]. MR resonance frequency is also dependent on the orientation of anisotropic structures in the magnetic field [2], which possibly contributes to deviations in the diffusion measurement. We investigated the effect of orientation of an anisotropic diffusion phantom on the non-DW (b0) signal, FA and the first eigen value (λ1). Our results show considerable variations in b0, FA and λ1 in the anisotropic phantom at different orientations in the magnetic field. Methods Acquisition A commercially available spherical diffusion phantom (Brain Innovation, NL)[3] with 3 tubular anisotropic phantoms in each principal direction was used. Fig 1 shows the opened physical phantom and a rendering of the phantoms x: red, y: green, z: blue. The sphere, marked with an angular grid was rotated in 3 planes in the scanner (x-z, y-z and x-y plane) in steps of ~45 degrees. At each orientation, a DW-MRI scan was made: Siemens Allegra 3T birdcage coil, FOV 192x192mm, 70 slices, 2x2x2 mm voxel size. b-value 1000 s/mm, 27 diffusion directions and 1 b0 image. A GRE phase and magnitude image were also obtained at the same resolution. TE(1)=10.0ms, TE(2)=12.46ms. Data Analysis Data analysis was done in FSL (FMRIB, UK). FA and the first eigenvector λ1 were calculated in each phantom orientation. The b0 image from each orientation was co-registered with the reference orientation (X, Y and Z phantoms aligned with x, y, z scanner axis) and the FA and λ1 images were transformed into the same principal orientation. The mean, re-oriented FA map was skeletonized to obtain ROIs containing voxels from the tubular phantoms for each principal direction. Results Fig 2 shows the b0 images of the phantom rotated in the x-z plane, with the angle of the z-phantom wrt the scanner’s zaxis. The phantom was rotated in a similar fashion in the other two planes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calculation of Positron Distribution in the Presence of a Uniform Magnetic Field for the Improvement of Positron Emission Tomography (PET) Imaging Using GEANT4 Toolkit

Introduction Range and diffusion of positron-emitting radiopharmaceuticals are important parameters for image resolution in positron emission tomography (PET). In this study, GEANT4 toolkit was applied to study positron diffusion in soft tissues with and without a magnetic field for six commonly used isotopes in PET imaging including 11C, 13N, 15O, 18F, 68Ga, and 82Rb. Materials and Methods GEA...

متن کامل

MRS Shimming: An Important Point Which Should not be Ignored

Introduction: Proton magnetic resonance spectroscopy (MRS) is a well-known device for analyzing the biological fluids metabolically. Obtaining accurate and reliable information via MRS needs a homogeneous magnetic field in order to provide well-defined peaks and uniform water suppression. There are lots of reasons which can disturb the magnetic field homogeneity which can be corrected by a proc...

متن کامل

Geometric distortion evaluation of magnetic resonance images by a new large field of view phantom for magnetic resonance based radiotherapy purposes

Background: The magnetic resonance imaging (MRI)-based radiotherapy planning method have been considered in recent years because of the advantages of MRI and the problems of planning with two images modality. The first step in MRI-based radiotherapy is to evaluate magnetic resonance (MR) images geometric distortion. Therefore, the present study aimed to evaluate system related geometric distort...

متن کامل

Fabrication of New 3D Phantom for the measurement of Geometric Distortion in Magnetic Resonance Imaging System

Introduction: Geometric distortion, an important parameter in neurology and oncology. The current study aimed to design and construct a new three-dimensional (3D) phantom using a 3D printer in order to measure geometric distortion and its 3D reproducibility. Material and Methods: In this study, a new phantom ...

متن کامل

Fabrication of New 3D Phantom for Measuring Geometric Distortion in Magnetic Resonance Imaging System

  Introduction: Geometric distortion is a major shortcoming of magnetic resonance imaging (MRI), which has an important influence on the accuracy of volumetric measurements, an important parameter in neurology and oncology. Our goal is to design and construct a new three- dimensional phantom using a 3D printer in order to measure geometric distortion and its reproducibility in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010