Performance of the coughassist insufflation-exsufflation device in the presence of an endotracheal tube or tracheostomy tube: a bench study.
نویسندگان
چکیده
BACKGROUND The CoughAssist is a mechanical insufflator-exsufflator designed to assist airway secretion clearance in patients with ineffective cough. The device may benefit intubated and tracheotomized patients. We assessed the impact of various artificial airways on peak expiratory flow (PEF) with the CoughAssist. METHODS We measured PEF and pressure at the airway opening in a lung model during insufflation-exsufflation with the CoughAssist, at 3 set pressures: 30/-30, 40/-40, and 50/-50 cm H(2)O, first without (control), and then with different sizes (6.5 to 8.5 mm inner diameter) of endotracheal tube (ETT) and tracheostomy tube (6, 7, and 8 mm inner diameter), compliance settings of 30 and 60 mL/cm H(2)O, and resistance settings of 0 and 5 cm H(2)O/L/s). We analyzed the relationship between PEF and pressure with linear regression. RESULTS With compliance of 30 mL/cm H(2)O and 0 resistance the slope of the control relationship between PEF and pressure was statistically significantly greater than during any conditions with ETT or tracheostomy tube. Therefore, in comparison to the control, the relationship of PEF to pressure significantly went in the direction from top to bottom as the ETT or tracheostomy tube became narrower. The findings were the same with compliance of 30 mL/cm H(2)O and resistance of 5 cm H(2)O/L/s. With compliance of 60 mL/cm H(2)O the highest set pressure values were not achieved and some relationships departed from linearity. The control slope of the relationship between PEF and pressure with compliance of 60 mL/cm H(2)O and 0 resistance did not significantly differ with any ETT or tracheostomy tube. CONCLUSIONS The artificial airways significantly reduced PEF during insufflation-exsufflation with CoughAssist; the narrower the inner diameter of the artificial airway, the lower the PEF for a given expiratory pressure.
منابع مشابه
Bench assessment of a new insufflation-exsufflation device.
BACKGROUND The Nippy Clearway is a new mechanical insufflation-exsufflation device used to assist cough. METHODS We compared the peak expiratory flow (PEF) with the Nippy Clearway versus the CoughAssist in a bench experiment. The relationship between PEF and pressure at the airway opening during exsufflation (minimum expiratory PAO) was investigated under 6 combinations of compliance (30 or 6...
متن کاملThe use of mechanical insufflation-exsufflation via artificial airways.
During the past 20 years, airway clearance techniques have been the subject of increasing scientific interest. Cough augmentation with mechanical insufflation-exsufflation is part of this trend. No fewer than 23 studies, of which 21 were published in the past 10 years, have improved the level of evidence on this technique.1-23 As suggested in Figure 1, the availability of insufflation-exsufflat...
متن کاملRepeated Tracheostomy Tube Cuff Rupture Due to Tracheobronchopathia Osteochondroplastica: A Case Report
Introduction: Tracheobronchopathia osteochondroplastica (TPO) is a rare benign disorder of the lower part of the trachea and the upper part of the main bronchi. Case Report: A case of tracheobronchopathia osteochondroplastica (TPO) diagnosed at the time of intubation in an intensive care unit due to difficulty when advancing the endotracheal tube beyond the vocal cords, is reported. A problem...
متن کاملDecanulation of patients with severe respiratory muscle insufficiency: efficacy of mechanical insufflation-exsufflation.
OBJECTIVE Ventilator dependent patients with neuromuscular disorders and high level spinal cord injury have been extubated and decanulated to continuous noninvasive intermittent positive pressure ventilatory support after mechanical insufflation-exsufflation was used to achieve specific criteria for tube removal. The purpose of this study is to report changes in extent of need for ventilator us...
متن کاملComparison of Alpha 200 and CoughAssist as intermittent positive pressure breathing devices: a bench study.
BACKGROUND Intermittent positive pressure breathing (IPPB) is used in non-intubated patients to increase lung volume and to enhance coughing. Alpha 200 (Salvia Lifetec, Kronberg, Germany) is a specific IPPB device. CoughAssist (Respironics France, Carquefou, France) is a mechanical insufflator-exsufflator used to remove secretions in patients with inefficient cough. Both can also be used for i...