Field study of lyophilised plasmas for local prothrombin time calibration in The Netherlands.
نویسنده
چکیده
AIM To assess the effect of a lyophilised calibrant plasma procedure on the international normalised ratio (INR) and its interlaboratory variation. METHODS INR equivalent values were assigned to five lyophilised plasmas (one from normal donors and four from coumarin treated patients) by a reference laboratory using three calibrated thromboplastin reagents. The calibrant plasmas and five artificial control blood specimens were mailed to 44 Dutch laboratories for prothrombin time (PT) determination. The assigned INR values were used to calculate calibration lines for each participant laboratory. The calibration lines were then used to translate the PT of the control specimens to INR. RESULTS For all lyophilised plasmas and control blood samples, there were significant differences between INR values determined with the three thromboplastin reagents. These differences could not be explained by inaccuracy of the international sensitivity index or mean normal PT of the reagents and must, therefore, have been induced by the preparation procedures for the lyophilised plasmas and control blood samples. The interlaboratory variation of the INR obtained with the calibrant plasma procedure had a coefficient of variation (CV) ranging between 2.1% and 7.3% and tended to be lower than the interlaboratory variation found with the usual methods (3.0-12.2% CV). There was a good agreement between the mean INRs obtained with the calibrant procedure and those obtained using the normal methods. CONCLUSIONS The present study highlights the limitations of some lyophilised plasmas and control blood samples. It is not possible to assign a single INR value to each of these lyophilised plasmas and control specimens that is valid for all thromboplastin reagents. Nevertheless, by using reagent specific INR equivalent values for the calibrant plasma procedure, the interlaboratory variation could be reduced.
منابع مشابه
European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors.
AIMS The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma...
متن کاملCalculation of System International Sensitivity Index: how many calibrant plasmas are required?
AIMS To determine the minimum number of calibrant plasmas required to assign an accurate System International Sensitivity Index (ISI). METHODS Eighteen lyophilised adsorbed and eight frozen coumarin plasmas were despatched to 57 participating laboratories. Each centre determined the prothrombin time of all of the plasmas using their routine system of coagulometer and thromboplastin reagent an...
متن کاملAssessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.
An attempt was made to correct for the effects of coagulometers on the International Sensitivity Index (ISI) in a series of collaborative studies. Modified ISI were derived from the prothrombin time results with coagulometer systems using a range of calibrated plasmas. Two alternative approaches to correction of the ISI were evaluated. The first relied on the consensus orthogonal regression slo...
متن کاملSimplified method for international normalized ratio (INR) derivation based on the prothrombin time/INR line: an international study.
BACKGROUND The need to perform local International Sensitivity Index (ISI) calibrations and in particular the requirement for a manual method for prothrombin time (PT) determination, have proved to be obstacles to application of the WHO scheme for PT standardization. METHODS We used international normalized ratio (INR) derived with a set of only 5 European Concerted Action on Anticoagulation ...
متن کاملGuidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determination--a rebuttal.
Reliable international normalized ratio (INR) determination depends on accurate values for international sensitivity index (ISI) and mean normal prothrombin time (MNPT). Local ISI calibration can be performed to obtain reliable INR. Alternatively, the laboratory may determine INR directly from a line relating local log(prothrombin time [PT]) to log(INR). This can be done by means of lyophilized...