Rapid-flow enthalpimetric determination of urea in serum, with use of an immobilized urease reactor.
نویسندگان
چکیده
We used a differential thermal detector in conjunction with an immobilized urease reactor to determine urea in serum. Samples (120 mul) are introduced into a flow stream and passed through an "adiabatic" column, which is packed with enough insolubilized urease to completely convert urea to ammonia and carbon dioxide. Measured temperature changes are directly proportional to the serum urea concentration. Urea in the presence of protein, bilirubin, and hemoglobin can thus be rapidly, simply, and inexpensively measured. Results correlate well with those obtained by the manual diacetyl monoxime and urease/indophenol methods.
منابع مشابه
Immobilized-enzyme nylon-tube reactors for routine determination of urea and citrulline in serum.
A continuous-flow clinical analyzer for the routine estimation of urea is described that makes use of an immobilized-enzyme nylon-tube reactor as part of a flow-through system (a Technicon AutoAnalyzer I). Results of blood-urea analyses by use of the immobilized urease are compared with determinations made with the diacetyl monoxime method and the urease solution method. Clinical trials carried...
متن کاملThe use of immobilized enzyme reactors in continuous-flow analyzers for the determination of glucose, urea, and uric acid.
We describe the use of immobilized enzymes in assay methods for the determination of glucose with glucose oxidase, uric acid with uricase, and urea with urease in serum samples. The enzyme reactor tubes were adapted to continuous-flow analyzers (Technicon AA II, SMA 12/60, and SMAC) used in routine laboratory determinations, and results with their use were compared to those from assays involvin...
متن کاملDetermination of urea in serum based on the combination of an enzymatic reaction with immobilized urease and ion chromatographic analysis.
A quantitative method for the determination of urea in serum was studied. An ion chromatograph (IC) with a conductivity detector was used in this method, where the chromatograph was modified by placing an immobilized urease column between the injection loop and a guard column of the cation analysis column. Immobilized urease was prepared by the adsorption of urease on cedar sawdust with triethy...
متن کاملA Conductometric Urea Biosensor by Direct Immobilization of Urease on Pt Electrode
Urease was immobilized on platinum electrode both by chemical binding and electropolymerization.The conductometric urea biosensor thus prepared showed a detection limit of 4.9×10-5 M and linear dynamic range from 4.9×10-5 to 5.8×10-3 M for urea concentration when the enzyme is covalently immobilized on Pt electrodes. Conductometric transducers respo...
متن کاملDetermination of urea in serum by using naturally immobilized urease in a flow injection conductimetric system.
A flow injection method was developed, aimed at the determination of urea in human serum. The system makes use of the naturally immobilized urease present in Canavalia ensiformis DC (jack bean). A column is filled with small pieces of this bean, and the sample (50 microliters) containing urea passes through it carried by a 1% NaCl solution. On leaving the column the stream is merged with an alk...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 22 8 شماره
صفحات -
تاریخ انتشار 1976