SU-E-J-100: Feasibility of Dose Calculation Using Combined Information of Cone-Beam and Multi-Slice CT Images.
نویسندگان
چکیده
PURPOSE The purpose of this study is to evaluate the feasibility of kilovoltage cone-beam CT (CBCT) images that are obtained with the Varian On-Board Imager in dose calculation at each radiation therapy. METHODS CBCT images are commonly degraded by scattered radiations originating in the patient's body, and so the CT numbers of the CBCT images depend on data acquisition conditions and the patient size. However, the anatomical shape of each organ is not likely affected by scattered radiations, and so we used only the shape of major organs such as lungs and bones in the CBCT images, and replaced these CT numbers with those of the multi-slice CT (MSCT) images that were used for dose calculation in a treatment planning. As regards this alternative CT number we adopted the median of MSCT numbers in a segmented region of a major organ each corresponding to that in the CBCT images. We evaluated the validity of our segmented region (SR) method with images of eight patients with lung diseases. The number of irradiation beams was four. In this evaluation we used the distance-to-agreement (DTA) and y analysis, and the dose-volume-histogram (DVH) analysis. RESULTS The pass rates of the DTA analysis (2mm) and γ analysis (2mm, 2%) between the dose distributions calculated with our method were 90.4±6.0% and 99.1±1.1%, respectively. The results of the DVH analysis showed that the differences in doses (average, maximum and minimum) for a target volume were 1.3±0.5%, 0.9±0.8% and 3.4±3.0%, respectively. These results showed that our method was acceptable in the calculation of a dose distribution. CONCLUSION We evaluated the dose calculation method with a combination of CBCT and MSCT images. This method could yield an accurate dose distribution and achieved an easier verification of radiation therapy on each treatment day.
منابع مشابه
Evaluation of MRI-based Polymer Gel Dosimetry for Measurement of CT Dose Index (CTDI) on 64 slices CT Scanners
Introduction: Computed tomography (CT) has numerous applications in clinical procedures but its main problem is its high radiation dose to the patients compared to other imaging modalities using x-ray. CT delivers approximately high doses to the nearby tissues due to the scattering effect, fan beam (beam divergence) and limited collimator efficiency. The radiation dose from multi-slice scanners...
متن کاملOptimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose
Introduction In computed tomography (CT) technology, an optimal radiation dose can be achieved via changing radiation parameters such as mA, pitch factor, rotation time and tube voltage (kVp) for diagnostic images. Materials and Methods In this study, the brain, abdomen, and thorax scaning was performed using Toshiba 16-slice scannerand standard AAPM and CTDI phantoms. AAPM phantom was used for...
متن کاملEffect of Slice Thickness on the Accuracy of Linear Measurements Made on Cone Beam Computed Tomography Images (InVitro
Objective: Cone beam computed tomography (CBCT) is applied for the imaging of the maxillofacial and dental structures, particularly for surgical treatments and dental implants. The aim of this study was to assess the effect of slice thickness on the accuracy of CBCT linear measurements. Methods: In this diagnostic accuracy study, forty-two titanium pins with the same dimensions were inserted...
متن کاملA feasibility study on the use of MV-CBCT images for urgent palliative treatment planning
Introduction: The application of 3D volumetric imaging modalities in treatment planning of radiation therapy can provide more precisely define tumor localization, and computed tomography (CT) is the most common accepted method for treatment planning. Given the lack of a CT scanner stationed in all radiotherapy departments and equipping most of the medical linear accelerators wi...
متن کاملThe Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 39 6Part7 شماره
صفحات -
تاریخ انتشار 2012