Evaluation of a point-of-care glucose and β-hydroxybutyrate meter operated in various environmental conditions in prepartum and postpartum sheep.

نویسندگان

  • Katlin J Hornig
  • Stacey R Byers
  • Robert J Callan
  • Timothy Holt
  • Megan Field
  • Hyungchul Han
چکیده

OBJECTIVE To compare β-hydroxybutyrate (BHB) and glucose concentrations measured with a dual-purpose point-of-care (POC) meter designed for use in humans and a laboratory biochemical analyzer (LBA) to determine whether the POC meter would be reliable for on-farm measurement of blood glucose and BHB concentrations in sheep in various environmental conditions and nutritional states. ANIMALS 36 pregnant mixed-breed ewes involved in a maternal feed restriction study. PROCEDURES Blood samples were collected from each sheep at multiple points throughout gestation and lactation to allow for tracking of gradually increasing metabolic hardship. Whole blood glucose and BHB concentrations were measured with the POC meter and compared with serum results obtained with an LBA. RESULTS 464 samples were collected. Whole blood BHB concentrations measured with the POC meter compared well with LBA results, and error grid analysis showed the POC values were acceptable. Whole blood glucose concentrations measured with the POC meter had more variation, compared with LBA values, over the glucose ranges evaluated. Results of error grid analysis of POC-measured glucose concentrations were not acceptable, indicating errors likely to result in needless treatment with glucose or other supplemental energy sources in normoglycemic sheep. CONCLUSIONS AND CLINICAL RELEVANCE The POC meter was user-friendly and performed well across a wide range of conditions. The meter was adequate for detection of pregnancy toxemia in sheep via whole blood BHB concentration. Results should be interpreted with caution when the POC meter is used to measure blood glucose concentrations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of energy related metabolites during peri-parturition period in single and twin-bearing Lori-Bakhtiari ewes

Background: The Lori-Bakhtiari sheep breed is one of the most common native breed in the south-western parts of Iran. The peri-parturition period (2 weeks before to, 2 weeks after parturition), is generally of critical importance regarding the health, production and profitability of the ewes. OBJECTIVES: This study was carried out to compare the serum concentrations of glucose, non-esterified f...

متن کامل

Measure of energy related biochemical metabolites changes during peri-partum period in Makouei breed sheep

Makouei sheep is one of the famous breeds in Iran which is reared in Azerbaijan province for their meat, milk and wool. Fifty clinically healthy Makouei ewes were selected to study the variations in energy-related blood metabolites during peri-partum period. Blood was collected from Jugular vein from each sheep on day 7 before the expected lambing time, day of parturition and also day 7 postpar...

متن کامل

Poly-β-hydroxybutyrate Production by Methylosinus trichosporium OB3b at Different Gas-phase Conditions

Background: The utilization of methane for production of Poly-β-hydroxybutyrate (PHB) not only cuts the emissions of greenhouse gases but also greatly reduces PHB production cost. Objectives: The aim of this study was to determine the effects of gas-phase conditions on PHB production by Methylosinus trichosporium OB3b. Ma...

متن کامل

Study of Glycation Process of Human Carbonic Anhydrase II and Investigation of Effect of Fasting On Enzyme Activity by Using Spectroscopic Methods

Background: Glycation is the non-enzymatic reaction between the carbonyl groups in sugar and free amino groups in proteins. this reaction leads to changes in structure and functions of proteins. Advanced glycation end products (AGEs) is the final stage in this process, which is highly oxidizing and destructive nature, causing many diabetic complications. Methods: In the present investigation, ...

متن کامل

Evaluation of a point-of-care electrochemical meter to detect subclinical ketosis and hypoglycaemia in lactating dairy cows.

OBJECTIVES To evaluate and validate a hand-held electrochemical meter (Precision Xtra®) as a screening test for subclinical ketosis and hypoglycaemia in lactating dairy cattle. DESIGN Method comparison study using a convenience sample. PROCEDURE Blood samples were collected into plain tubes from the coccygeal vessels of 181 Holstein cows at 2-4 weeks of lactation during summer in Iran. Bloo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • American journal of veterinary research

دوره 74 8  شماره 

صفحات  -

تاریخ انتشار 2013