Bench study of automated tube compensation versus pressure support after extubation in icu patients

نویسندگان

  • L Baboi
  • H Penet
  • A Stoian
  • H Yonis
  • F Gobert
  • R Tapponnier
  • F Bayle
  • V Leray
  • J-C Richard
  • C Guérin
چکیده

Methods Once removed the endotracheal tube from patients, who were extubated in our ICU after successful weaning trial, was connected to a filter (Hygrobac), and both were attached to ASL 5000 active servo lung (IngMar medical). The lung model was set with a compliance of 50 ml/ cmH2O and resistance of 10 cm H2O/L/s. The inspiratory effort included sinusoidal half-wave shape, muscular pressure -10 cmH2O, respiratory rate 27 breaths/min. The ventilator used was Evita XL (Dräger) set as used in patients during the weaning trial in our ICU, ie pressure support 7 cm H2O above PEEP of 4 cm H2O. ATC was set to 80% compensation and the internal diameter of the endotracheal tube for each patient was entered into the ventilator dialog box. Pressure support was run for 2 minutes without then with ATC. Pressure time product (PTP) trigger was defined as the product of airway pressure to time spent between onset of inspiratory effort and pressurization. The quality of airway pressurization was assessed as the area of airway pressure over time from inspiratory effort to 0.5 s later (PTP500). PTP500 was further expressed as percent of ideal PTP500, which was airway pressure recorded at the end of inspiratory time multiplied by 0.5 s. The values are expressed as median [first-third quartiles]. The comparisons were made by Wilcoxon signed rank test with continuity correction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Automatic Tube Compensation versus Pressure Support Ventilation and Extubation Outcome in Children: A Randomized Controlled Study

Background. Automatic tube compensation (ATC) has been developed to overcome the imposed work of breathing due to artificial airways during spontaneous breathing trials (SBTs). Objectives. This study aimed to assess extubation outcome after an SBT (spontaneous breathing trial) with ATC compared with pressure support ventilation (PSV) and to determine the risk factors for extubation failure. Met...

متن کامل

The effect of dexmedetomidine on the hemodynamic responses of endotracheal extubation and sedation level in recovery after cholecystectomy

Background: Emergence from general anesthesia and removing of tracheal tube can be associated with coughing, agitation, and hemodynamic disturbances. Dexmedetomidine is an alpha two adrenoceptor agonist that has effective sedation with less cardiovascular unstability and respiratory depression and may be useful for extubation and prevention of hemodynamic response during tracheal tube removing....

متن کامل

The Comparison of Automatic Tube Compensation (ATC) and T-piece During Weaning.

OBJECTIVE Automatic Tube Compensation (ATC) is a newly developed mechanical ventilatory support method. The aim of this study was to compare the ATC and the T-piece as a weaning method. METHODS Patients who were treated in ICU with mechanical ventilation for longer than 24 hours were included in this prospective clinical study. Fifty patients were divided into two groups for weaning, ATC or T...

متن کامل

Comparison of Extubation of Patients in Prone versus Supine Position in workers Undergoing Lumbar Disc Surgery

Introduction: This study compared the extubation of patients in prone versus supine positions in patients undergoing lumbar disc surgery. Methods: In this single-blind clinical trial, 60 patients who were candidates for elective lumbar disc surgery were selected and randomly assigned into two groups of 30 patients. All patients primarily underwent surgery in the prone position. After surgery a...

متن کامل

Effect of Stabilizing Endotracheal Tube on Quality of Airway Management in Patients Hospitalized in ICU

Background: The unplanned departure of endotracheal tube is one of the main complications for endotracheal intubation. In addition to endotracheal extubation, its movement can also lead to damage to the oral cavity. Stabilizing endotracheal tube is the most important factor that can prevent unplanned departure of endotracheal tube as well as its movement. The current study was done with the aim...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2015