Acetabular reconstruction using a Kerboull cross-plate, structural allograft and cemented dual-mobility cup in revision THA at a minimum 5-year follow-up.
نویسندگان
چکیده
The current study aimed to evaluate the outcome of a continuous and prospective series of 61 revision THAs with AAOS grade III and IV acetabular bone defect reconstruction using a Kerboull cross-plate, structural allograft and cemented dual mobility cup (Saturne, Amplitude, Valence, France). At a 7.5-year mean follow-up, no instability was reported after revision. In addition, no failure of the acetabular reconstruction was observed in 98% of the patients with complete allograft osseointegration and no evidence of mechanical rupture of the Kerboull cross-plate and/or loosening of the cemented dual mobility cup. In conclusion, such reconstruction technique demonstrated excellent results at mid-term follow-up in terms of prevention of instability after revision, restoration of the acetabular bone stock, and stable cemented fixation of the dual mobility cup.
منابع مشابه
Risk factors for failure of revision total hip arthroplasty using a Kerboull-type acetabular reinforcement device
BACKGROUND The present study aimed to identify the risk factors associated with revision total hip arthroplasty (THA) failure using a Kerboull-type (KT) plate. METHODS We analyzed 77 revision THAs using cemented acetabular components with a KT plate for aseptic loosening between May 2000 and March 2012. We examined the association of bone graft type, acetabular bone defects, age at the time o...
متن کاملStructural and Morselized Allografting Combined with a Cementless Cup for Acetabular Defects in Revision Total Hip Arthroplasty: A 4- to 14-Year Follow-Up
Using morselized and structural allograft to restore bone stock for massive acetabular bone defect in revision total hip arthroplasty (THA) is an appealing procedure. However, concerns about inability to achieve long-term stability following allograft resorption remained. From 2003 to 2012, 59 hips in 58 patients undergoing revision THA for Paprosky type II or III acetabular defects were retros...
متن کاملDoes a Cemented Cage Improve Revision THA for Severe Acetabular Defects?
BACKGROUND Evidence suggests a growing incidence of revision total hip arthroplasty (THA) including a subset with large acetabular defects. Revision THA for severe acetabular bone loss is associated with a relatively high rate of mechanical failure. QUESTIONS/PURPOSES We questioned whether cementing a cage to the reconstructed acetabular defect and pelvis would improve the rate of mechanical ...
متن کاملImpaction grafting in association with the Charnley-Kerboull cemented femoral component: operative technique and two- to 16-year follow-up results.
We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30. There was one intra-operative fracture of the femur but none later. Two ...
متن کاملBone Remodeling in Acetabular Reconstruction Using a Kerboull-Type Reinforcement Device and Structural Bone-Grafting in Total Hip Arthroplasty.
BACKGROUND The purpose of this study was to identify the long-term durability of the Kerboull-type reinforcement device (KT plate) in acetabular reconstruction for massive bone defects, assessing the remodeling of structural bone grafts. METHODS This study retrospectively evaluated 106 hips that underwent acetabular reconstruction using a KT plate between November 2000 and December 2010. Thir...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of arthroplasty
دوره 29 2 شماره
صفحات -
تاریخ انتشار 2014