Performance of the Du Pont aca ammonia method.

نویسندگان

  • B T Doumas
  • L L Hause
  • R D Sciacca
  • B Jendrzejczak
  • C C Foreback
  • J D Hoover
  • W W Spencer
  • P L Smock
چکیده

We evaluated the performance of the Du Pont aca ammonia procedure with regard to (a) linearity, (b) precision, (c) interferences, and (d) effect of anticoagulants. Linearity extends to 2,000 mumol/L. The SD of the method was essentially constant (2 to 3 mumol/L) and independent of the NH3 concentration. Hemoglobin, bilirubin, and lipemia do not interfere. Either EDTA or heparin is suitable as anticoagulant. Recovery of NH3 added to plasma samples averaged 102% (range: 97--107%). We established normal values by measuring NH3 in 188 plasma samples from apparently healthy individuals. The 95% confidence range is from 10 to 35 mumol/L. The aca ammonia method compares very well with that of Kingsley and Tager but correlates less strongly with that of Reinhold and Chung. We describe a protein-based solution with stable NH3 concentration that is suitable as a control material.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An alternative method for assaying cerebrospinal fluid protein in the presence of methotrexate.

Therapeutic concentrations of methotrexate can cause significant positive interference in cerebrospinal fluid (CSF) protein values when assayed in the Du Pont aca. Conversely, our modified turbidimetric method, in which trichloroacetic acid (TCA) plus a sample blank containing dilute hydrochloric acid is used in place of TCA, exhibits little or no interference from methotrexate. This was verifi...

متن کامل

Negative interference with the Du Pont aca method for measuring digoxin.

In response to a recent Letter (1) concerning the digoxin method in the aca Discrete Clinical Analyzer, the Dii Pont Company would like to offer the following. As in any enzyme immunoassay, the possibility existsthattheenzyme could be either activated or inhibited by drugs and (or) nonspecific substances, resulting in reported values that do not reflect true concentrations of the drug. We have ...

متن کامل

Albumin activation of urinary amylase as determined with the Du Pont aca.

Protein activation of urinary alpha-amylase (EC 3.2.1.1) activity was observed during an evaluation of the Du Pont aca procedure for the determination of urinary alpha-amylase. This activation effect became constant for urinary albumin concentrations exceeding 1.50 g/liter. It is recommended that urinary alpha-amylase be analyzed with sufficient albumin added to maximize this effect. The aca al...

متن کامل

Automated enzymatic assay for plasma ammonia.

An enzymatic method for determining plasma ammonia with the Du Pont Automatic Clinical Analyzer (aca) is described. The assay requires a sample volume of 500 muL for a kinetic ammonia measurement. The reaction is initiated with glutamate dehydrogenase and the rate of depletion of NADPH is monitored with two measurements, 17 s apart, at 340 nm. Reaction conditions have been optimized for maximum...

متن کامل

Salicylate determined with a microcentrifugal analyzer, and compared with Du Pont aca, trinder, and liquid-chromatographic methods.

We describe a new automated method for measuring serum salicylate in the Multistat III microcentrifugal analyzer. Ferric nitrate reagent and serum blanking are used. We compare this new method, the automated Du Pont aca method, and the manual Trinder method with a "high-performance" liquid-chromatographic method. The unblanked Trinder method had the poorest correlation (r = 0.980, Sy X x = 19.1...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 25 1  شماره 

صفحات  -

تاریخ انتشار 1979