Strut fracture of Björk-Shiley convexo-concave valve in Japan--risk of small valve size--.

نویسندگان

  • S Watarida
  • S Shiraishi
  • T Nishi
  • M Imura
  • Y Yamamoto
  • R Hirokawa
  • M Fujita
چکیده

BACKGROUND The Björk-Shiley convexo-concave (BSCC) prosthetic valve was introduced in 1979. Between 1979 and 1986, approximately 86,000 BSCC valves were implanted. By December 31, 1994, 564 complete strut fractures had been reported to the manufacture. We experienced a case of an outlet strut fracture and investigated the risk of BSCC prosthetic valve fractures in Japan. METHODS AND RESULTS To investigate the risk factor of a strut fracture in Japan, we investigated published cases of strut fractures. Between 1979 and 1986, 2021 BSCC valves were implanted in Japan. By January 31, 2000, 11 complete strut fractures of 60-degree BSCC valves including our case had occurred. The patients were eight males and three females. The average age at valve replacement was 42.4+/-8.1 years, and nine of eleven (81.8%) were patients < 50 years-old. The average age of the patients when the BSCC valve fractured was 47.7+/-6.4 years, and eight of eleven (72.7%) were patients <= 50 years old. All patients were implanted in the mitral position. The sizes of the BSCC valve were 27 mm (n=5) (45.5%), 29 mm (n=3) (27.2%), and 31 mm (n=3) (27.2%). Four patients died and seven patients survived. CONCLUSIONS Although only 11 BSCC valve struts fractured and statistical analysis could not be performed, our findings suggest that the high risk group for a strut fracture in Japan is young male patients with a mitral valve, >= 27 mm in size with BSCC models manufactured before March 1982. When following-up patients with BSCC models manufactured before March 1982, the possibility of a strut fracture in all BSCC valve sizes should be kept in mind.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Prophylactic replacement of Bjŏrk-Shiley convexo-concave heart valves: an easy-to-use tool to aid decision-making in individual patients.

OBJECTIVE To develop an easy-to-use tool for decision-making on prophylactic replacement of Björk-Shiley convexo-concave heart valves. DESIGN Decision analysis comparing elective replacement with observation. SETTING Referral centres for patients with artificial heart valves. PATIENTS Quantitative estimates were obtained from a follow up study conducted in The Netherlands, including 2303 ...

متن کامل

Psychological distress among recipients of Björk-Shiley convexo-concave valves: the impact of information.

OBJECTIVE To establish whether there is more psychological distress among recipients of Björk-Shiley convexo-concave (BScc) valves than among recipients of other valves not known to fracture. DESIGN Cross sectional study. PATIENTS 137 patients who received either a BScc valve or a Sorin Biomedical spherical valve during a randomised trial between 1982 and 1983 at St Antonius Hospital, Nieuw...

متن کامل

The Björk-Shiley convexo-concave heart valve experience from the perspective of the supervisory panel.

The 20-year activities of a medical supervisory panel appointed under the terms of a settlement agreement of the Bowling v. Pfizer class action suit involving the Björk-Shiley convexo-concave (BSCC) heart valve are detailed. Of approximately 86,000 valves implanted, catastrophic failure of the valve was reported in 663 patients from 1978 to 2012. In 1994, a 7-member medical panel consisting of ...

متن کامل

Age thresholds for prophylactic replacement of Björk-Shiley convexo-concave heart valves. A clinical and economic evaluation.

BACKGROUND Björk-Shiley convexo-concave heart valves have an increased risk of mechanical failure. One might consider prophylactic rereplacement as a preventive measure to avert the disastrous consequences of these failures. We investigated the effect that prophylactic rereplacement has on survival of individual patients and on the medical costs. METHODS AND RESULTS Quantitative estimates for...

متن کامل

Decision guidelines for prophylactic replacement of Björk-Shiley convexo-concave heart valves: impact on clinical practice.

BACKGROUND Because of risk of outlet strut fracture, prophylactic replacement should be considered for Björk-Shiley convexo-concave (BScc) valve recipients. We assessed the effects of epidemiological and decision-analytic guidelines on actual BScc valve replacement. METHODS AND RESULTS We performed a retrospective cohort study including all 2263 Dutch BScc patients with a mean follow-up of 11...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia

دوره 7 4  شماره 

صفحات  -

تاریخ انتشار 2001