Comparison of ESI– and APCI–LC–MS/MS methods: A case study of levonorgestrel in human plasma☆
نویسندگان
چکیده
Electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) techniques for liquid chromatography-tandem mass spectrometry (LC-MS/MS) determination of levonorgestrel were evaluated. In consideration of difference in ionization mechanism, the two ionization sources were compared in terms of LC conditions, MS parameters and performance of method. The sensitivity for detection of levonorgestrel with ESI was 0.25 ng/mL which was lower than 1 ng/mL with APCI. Matrix effects were evaluated for levonorgestrel and canrenone (internal standard, IS) in human plasma, and the results showed that APCI source appeared to be slightly less liable to matrix effect than ESI source. With an overall consideration, ESI was chosen as a better ionization technique for rapid and sensitive quantification of levonorgestrel. The optimized LC-ESI-MS/MS method was validated for a linear range of 0.25-50 ng/mL with a correlation coefficient ≥0.99. The intra- and inter-batch precision and accuracy were within 11.72% and 6.58%, respectively. The application of this method was demonstrated by a bioequivalence study following a single oral administration of 1.5 mg levonorgestrel tablets in 21 Chinese healthy female volunteers.
منابع مشابه
A quick and Sensitive Liquid Chromatography–tandem Mass Spectrometry (LC-MS) Method for the Determination of Enalapril and Enalaprilat in Human Plasma: Application to a Bioequivalence Study
A rapid and sensitive liquid chromatography–tandem mass spectrometry (LC-MS) method for the estimation of enalapril and enalaprilat in human plasma. Detection of analytes was achieved by tandem mass spectrometry with electrospray ionization (ESI) interface in positive ion mode was operated under the multiple-reaction monitoring mode. Sample pretreatment involved in a one-step protein precipitat...
متن کاملLiquid chromatography–tandem mass spectrometry (LC-MS) method for the assignment of enalapril and enalaprilat in human plasma
A rapid and sensitive liquid chromatography–tandem mass spectrometry (LC-MS) method was developed for the determination of enalapril and enalaprilat in human plasma. Detection of analytes was achieved by tandem mass spectrometry with electrospray ionization (ESI) interface in positive ion mode which was operated under the multiple-reaction monitoring mode. Sample pretreatment was involved...
متن کاملMethod validation of clonidine hydrochloride in human plasma by LC-MS technique
A simple and sensitive high performance liquid chromatography-electrospray ionization mass spectrometry method has been evaluated for the assignment of clonidine hydrochloride in human plasma. The mobile phase composed of acetonitrile–water 60:40 (v/v), and 0.2% formic acid 20 µL of sample was chromatographically analyzed using a repacked ZORBAX-XDB-ODS C18 column (2.1 mm×30 m...
متن کاملMatrix effect in bio-analysis of illicit drugs with LC-MS/MS: influence of ionization type, sample preparation, and biofluid.
The purpose of the present work was to evaluate the synergistic effect of ionization type, sample preparation technique, and bio-fluid on the presence of matrix effect in quantitative liquid chromatography (LC)-MS/MS analysis of illicit drugs by post-column infusion experiments with morphine (10-microg/mL solution). Three bio-fluids (urine, oral fluid, and plasma) were pretreated with four samp...
متن کاملMethod validation of methotrexate in human plasma by LC-MS technique in patients with brain tumor.
Background and aims: A selective and sensitive high performance liquid chromatography-electrospray ionization mass spectrometry method has been established for determination of methotrexate in human plasma. Methods: Methotrexate was extracted from plasma with acetonitrile. The mobile phase consisted of acetonitrile-water-formic acid 74: 25: 1(v/v). Twenty...
متن کامل