Comparison of the Vitek gram-positive susceptibility 106 card, the MRSA-Screen latex agglutination test, and mecA analysis for detecting oxacillin resistance in a geographically diverse collection of clinical isolates of coagulase-negative staphylococci.

نویسندگان

  • T Yamazumi
  • I Furuta
  • D J Diekema
  • M A Pfaller
  • R N Jones
چکیده

The Vitek automated susceptibility testing system with a modified gram-positive susceptibility (GPS) 106 card (bioMerieux Vitek, Inc., Hazelwood. Mo.) and a rapid slide latex agglutination test (MRSA-Screen test; Denka Seiken Co., Ltd., Tokyo, Japan) were evaluated for their abilities to detect oxacillin resistance in coagulase-negative staphylococci (CoNS). The reference broth microdilution method and the detection of the mecA gene by PCR ("gold standard" reference result) were used to compare the results obtained with the commercial products. A total of 123 clinical isolates consisting of eight species were selected from U.S. surveillance collections. Among the mecA-positive isolates (95 strains), 30 isolates were initially negative on the MRSA-Screen test read at 3 min. When the agglutination reaction was extended for 10 min, 26 of the 30 isolates became positive. For a different four isolates, the oxacillin MIC was < or =0.25 microg/ml on the Vitek GPS 106 card. Among the mecA-negative isolates (28 strains), for two Staphylococcus warneri, two S. lugdunensis, and two S. saprophyticus strains MICs were > or =0.5 microg/ml by the reference broth microdilution method. Four of these isolates were also categorized as resistant with the Vitek GPS 106 card and two isolates were positive by the MRSA-Screen test. Overall, the MRSA-Screen test, GPS 106 card, and reference broth microdilution method had sensitivities of 95.7 (result at 10 min), 95.7, and 100%, respectively, and specificities of 92.8, 85.7, and 78.5%, respectively. Although the MRSA-Screen test required a slight procedural modification, both commercial methods achieved a sensitivity and specificity at detecting oxacillin resistance in CoNS at a level that was acceptable for clinical laboratory use.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the methicillin-resistant Staphylococcus aureus (MRSA)-Screen latex agglutination test for detection of MRSA of animal origin.

Methicillin (oxacillin)-resistant staphylococci (MRS) have emerged as major clinical and epidemiological pathogens, and there have been frequent reports of MRS infections in the veterinary field. The MRSA-Screen latex agglutination test (Denka Seiken Co., Ltd., Tokyo, Japan) was compared with an oxacillin agar screen test, MIC determination, and mecA PCR assay, the "gold standard." In an analys...

متن کامل

Rapid detection of mecA-positive and mecA-negative coagulase-negative staphylococci by an anti-penicillin binding protein 2a slide latex agglutination test.

A rapid slide latex agglutination (LA) test, MRSA-Screen (Denka Seiken Co., Niigata, Japan), which detects PBP 2a, was tested for its ability to differentiate between mecA-positive and -negative coagulase-negative staphylococci. A total of 463 isolates from 13 species were included in the study. The mecA gene was detected by PCR, and the oxacillin MIC was determined by the agar dilution method ...

متن کامل

Laboratory detection methods for methicillin resistance in coagulase negative Staphylococcus isolated from ophthalmic infections.

PURPOSE To evaluate different methods of oxacillin susceptibility testing of ocular isolates, considering polymerase chain reaction (PCR) as the 'gold standard', and to compare the in vitro susceptibility to oxacillin with that of other antimicrobials used in ophthalmologic practice. METHODS The Vitek gram-positive identification card was used to identify ocular coagulase negative Staphylococ...

متن کامل

Evaluation of the vitek 2 AST-P559 card for detection of oxacillin resistance in Staphylococcus aureus.

Staphylococcus aureus strains with resistance to methicillin or oxacillin (MRSA) represent one of the main nosocomial pathogens at present. MRSA infections are clearly associated with higher mortality and economic cost than those caused by methicillin-susceptible S. aureus (1). In Spain, the prevalence of methicillin resistance among S. aureus isolates has increased since the 1990s and in some ...

متن کامل

Rapid detection of methicillin resistance in Staphylococcus aureus isolates by the MRSA-screen latex agglutination test.

The slide agglutination test MRSA-Screen (Denka Seiken Co., Niigata, Japan) was compared with the mecA PCR ("gold standard") for the detection of methicillin resistance in Staphylococcus aureus. The MRSA-Screen test detected the penicillin-binding protein 2a (PBP2a) antigen in 87 of 90 genetically diverse methicillin-resistant S. aureus (MRSA) stock culture strains, leading to a sensitivity of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 39 10  شماره 

صفحات  -

تاریخ انتشار 2001