Improved Self-Referenced Parallel MRI imaging in EPI by using UNFOLD to Remove Nyquist Ghosts

نویسندگان

  • W. S. Hoge
  • H. Tan
  • R. A. Kraft
چکیده

Introduction: The quality of parallel MRI (pMRI) reconstructions is directly dependent on the data used for coil sensitivity estimation in SENSE [1] and/or GRAPPA [2] reconstruction coefficient calibration. In many applications, it is desirable to use self-referenced pMRI approaches over a prescan approach, to limit the effect of changes (such as motion) that occur between acquisition of the prescan pMRI calibration data and the accelerated image data. In EPI acquisitions, self-referenced approaches are confounded by the presence of Nyquist ghosts, which are a persistent artifact in EPI. These artifacts occur when data sampled along positive and negative read-out gradients is inconsistent. A well-accepted way, [3], to correct for Nyquist ghost artifacts is to measure the shift in k-space between positive and negative readout data and realign the two data sets before reconstructing the images. Unfortunately, this approach also commonly uses reference scan data. More recently, methods such as 2D phase-mapping [4] and phase labeling (PLACE) [5] have been proposed to remove Nyquist ghosts. These methods use a slightly modified EPI k-space trajectory and then interleave the data from two frames to produce one image data set acquired only on positive read-outs and a second image data set acquired on only negative readouts. While each separate image is free of Nyquist ghosts, the disadvantage is that interleaving reduces the temporal sampling rate by half. We propose an approach that employs UNFOLD [6] in place of interleaving. This approach effectively eliminates Nyquist ghosts, which enables self-referenced parallel imaging while simultaneously maintaining a high temporal imaging rate. Theory: Employing a PLACE sampling trajectory modulation, we propose to use UNFOLD instead of interleaving to form images that satisfy the Nyquist sampling criterion. UNFOLD uses temporal encoding to alternate the polarity of aliasing artifacts in a temporal sequence. That is, a shift in the sampling grid (1Δk) will introduce a phase change in the image domain. This induces aliasing artifacts which result from sub-sampling to alternate in polarity from frame to frame—e.g. to ’flicker’ at the temporal Nyquist rate. These temporal encoded aliasing artifacts can then be subsequently removed using a narrow high-frequency notch filter in the temporal dimension. Consider that shifting the EPI data sampling grid by (1Δk), as in PLACE, introduces a temporal modulation. Separating the data into two sets—one for a positive readouts and the other for negative read-outs—will result in the sub-sampling commonly used with UNFOLD. Each set can then be processed separately using a temporal UNFOLD filter, and then combined to produce an image series free of Nyquist ghosts. Methods: Data was acquired on a GE Signa 1.5T scanner using a standard 8-channel head coil. The EPI trajectory was modified to shift the sampling grid on alternate frames by (1Δk) along ky. Further, the EPI blip table was modified to provide variable density sub-sampling, acquiring 8 ky lines (1Δk) apart near the center of k-space, and gradually increasing the gap between ky lines to (5Δk) in the high frequency region. An acceleration factor of 3.2x was used, producing 40 ky lines for the 128x128 images acquired. A self-referenced double-pass (DP) GRAPPA algorithm with a 2x5 kernel was used to reconstruct the images [7]. On the first pass, data in the (2Δk) region was reconstructed using 8 central ACS lines for self-calibration. On the second pass, 16 ACS lines from the output of the first pass was used for self-calibration to reconstruct the remaining k-space data. Results: Fig 1 shows images of a water phantom reconstructed from (a) self-referenced DP GRAPPA with Nyquist ghost correction as in [3], (b) GRAPPA using 20 lines from a reference frame with Nyquist ghost correction as in [3], and (c) selfreferenced DP GRAPPA using UNFOLD for Nyquist ghost correction. The image in (c) demonstrates that using UNFOLD to remove Nyquist ghosts yields a significant improvement in image quality over previous methods, enabling self-referenced pMRI in EPI with significantly lower artifact levels as shown by the arrows. Discussion: The ability to successfully reconstruct selfreferenced pMRI images in EPI—free of Nyquist ghosts—opens many opportunities, such as the scenario shown in Fig 2, where three 3.2x self-referenced pMRI images can be acquired after a single excitation, by sampling through 3 sweeps of k-space in quick succession---demonstrating the transition from T1 to T2 weighting as the time from excitation grows longer. Other applications such as accelerated real-time fMRI can now be considered. The cost of using UNFOLD is a small loss in the temporal bandwidth (~ 10%) which depends on the bandwidth of the aliasing artifact suppression filter.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Method to Remove Nyquist Ghosts from Echo Planar Images (EPI) using UNFOLD

Introduction: Nyquist ghosts are a persistent artifact in echo planar imaging (EPI), and occur when data sampled along positive readout gradients and negative read-out gradients is inconsistent. There are many approaches to correct such artifacts including double-sampled (DS) EPI [1], 1D phase-correction with an algorithm developed by Ahn and Cho (AC) [2], 2D phase-mapping [3], and phase labeli...

متن کامل

High-resolution fMRI with higher-order generalized series imaging and parallel imaging techniques (HGS-parallel).

PURPOSE To develop a novel approach for high-resolution functional MRI (fMRI) using the conventional gradient-echo sequence. MATERIALS AND METHODS Echo-planar imaging (EPI) techniques have generally been used for fMRI studies due to their fast imaging time. However, it is difficult for studying brain function at the submillimeter level using this sequence. In addition, EPI techniques have som...

متن کامل

Robust Elimination of EPI Nyquist Ghosts via Spatial and Temporal Encoding

Introduction: Nyquist ghosts are an inherent artifact in echo planar imaging (EPI). Many methods have been proposed to correct for them, with various strengths and weaknesses. Here, we propose a fusion of previous spatial [1] and temporal [2] encoding methods to achieve the goal of self-referenced ghost elimination with very low temporal latency. Methods: Our approach follows from the method de...

متن کامل

A Generalized Phase Correction Technique for EPI-PROPELLER

INTRODUCTION: PROPELLER (Periodically rotated overlapping parallel lines with enhanced reconstruction [1]) based on echo planar imaging (EPI) [2,3] can acquire wider blades with increased time efficiency compared to fast-spin echo (FSE)-based PROPELLER [4]. However, EPI-based sequences suffer from Nyquist ghosts arising primarily from eddy currents and gradient anisotropy [5], thus requiring ph...

متن کامل

Robust EPI Nyquist ghost elimination via spatial and temporal encoding.

Nyquist ghosts are an inherent artifact in echo planar imaging acquisitions. An approach to robustly eliminate Nyquist ghosts is presented that integrates two previous Nyquist ghost correction techniques: temporal domain encoding (phase labeling for additional coordinate encoding: PLACE and spatial domain encoding (phased array ghost elimination: PAGE). Temporal encoding modulates the echo plan...

متن کامل