Dependence of RF heating on SAR and implant position in a 1.5T MR system.
نویسندگان
چکیده
PURPOSE We evaluated radiofrequency (RF) heating of a humerus implant embedded in a gel phantom during magnetic resonance (MR) imaging for the specific absorption rate (SAR), angle between the implant and static magnetic field (B(0)), and position of the implant in the irradiation coil. METHODS We embedded a stainless steel humerus implant 2 cm deep in tissue-equivalent loop and mass phantoms, placed it parallel to the static magnetic field of a 1.5T MR scanner, and recorded the temperatures of the implant surface with RF-transparent fiberoptic sensors. We measured rises in temperature at the tips of the implant by varying the SAR from 0.2 to 4.0 W/kg and evaluated RF heating of the implant for its angle to B(0) and its displacement along B(0) from the center of the RF irradiation coil. RESULTS RF heating was similar for the loop and mass phantoms because the eddy current flows through the periphery of both. As the SAR increased, the temperature at the implant tip increased, and there was a linear relationship between the SAR and temperature rise. The values were 6.4 degrees C at 2.0 W/kg and 12.7 degrees C at 4.0 W/kg. Rise in temperature decreased steeply as the angle between the implant and B(0) surpassed 45 degrees . In addition, as the implant was displaced from the center of the RF coil to both ends, the rise in temperature decreased. CONCLUSION The rise in temperature in deep tissue was estimated to be higher than 1.0 degrees C for SAR above 0.4 W/kg. RF heating was greatest when the implant was set parallel to B(0). In MR imaging of patients with implants, there is a risk of RF heating when the loop of the eddy current is formed inside the body.
منابع مشابه
Parallel Excitation in Ultrahigh Field Human MR Imaging and Multi-Channel Transmit System.
Parallel excitation [1,2] with a multi-element Radio Frequency (RF) transceiver array [3-9] as a contemporary methodology has been advocated for human MR imaging at ultrahigh magnetic fields (7 Tesla and above). In ultrahigh field MRI, the required high operating frequency and thus shortened wavelength of radio frequency waves creates a complex wave behavior and increased phase variation of RF ...
متن کاملDesign and Evaluation of a Hybrid Radiofrequency Applicator for Magnetic Resonance Imaging and RF Induced Hyperthermia: Electromagnetic Field Simulations up to 14.0 Tesla and Proof-of-Concept at 7.0 Tesla
This work demonstrates the feasibility of a hybrid radiofrequency (RF) applicator that supports magnetic resonance (MR) imaging and MR controlled targeted RF heating at ultrahigh magnetic fields (B0≥7.0T). For this purpose a virtual and an experimental configuration of an 8-channel transmit/receive (TX/RX) hybrid RF applicator was designed. For TX/RX bow tie antenna electric dipoles were employ...
متن کاملSafety of metallic implants in magnetic resonance imaging.
Magnetic resonance (MR) imaging has become a commonly accepted medical procedure. Manufacturers of medical implant devices are submitting claims that their devices are safe and effective in a MR environment. This paper concentrates on the issue of heating of patients due to the interaction of metallic implants with the strong radiofrequency (RF) magnetic field produced by the MR scanner. The co...
متن کاملMR safety assessment of potential RF heating from cranial fixation plates at 7 T.
PURPOSE The increasing number of clinically oriented MRI studies at 7 T motivates the safety assessment of implants, since many 7 T research sites conservatively exclude all subjects with metallic implants, regardless of type or location. The purpose of this study was to investigate potential RF-induced heating during a 7 T MRI scan using a self-built transmit/receive RF coil in patients with i...
متن کاملContinuous Monitoring of RF-safety for Implantable MR-conditional Devices
Introduction The presence of an implantable active device (e.g. pacemaker or deep brain stimulator) usually represents a contraindication for an MR examination. Currently, effort is spent on the development of MR-conditional implantables. An important safety risk factor of all active implantable devices is potential RF coupling to conducting parts which may lead to RF heating. RF coupling is a ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine
دوره 6 4 شماره
صفحات -
تاریخ انتشار 2007