The Development of a Column-Switching High-Performance Liquid Chromatograph System for Chiral Separation of Mouse Urinary D, L-Lactate
نویسندگان
چکیده
Two high-performance liquid chromatograph (HPLC) systems with a Rheodyne Model 7000 switching valve for separation of Dand L-lactate in biological samples has been reported. However, since the switching valve was rotated manually, so that only 10 samples could be determined within one day. For better efficiency, a circuit for rotating the switching valve automatically, which is including mechanical units of motion, two timers, and electronic control units was designed. Total (D+L)-lactate fluorescent derivatives was separated from others by a TSKgel ODS-80Ts column, when the peak of lactate derivatives appeared in the chromatogram on the ODS column, the valve position was changed by switching and then the isolated lactate derivatives was introduced into the chiral column automatically. Then the Dand Lenantiomers were separated by the chiral column. The utility of the analyzer system was tested for the determination of chiral separation of urinary D, L-lactate in the mouse. It takes about 90 min to analyze one urine sample for a circulation. Before automation, we must turn the switching valve manually when the signal shows up, so that there must be someone always standby and only about 10 samples could be analyzed within one day. With the designed automatic column-switching HPLC system, the switching valve can be automatically turned at the setting time repeatedly and over 15 samples could be determined within one day. The automatic system makes the analysis of D-lactate be timesaving, convenient and reproducible.
منابع مشابه
Development of a new method based on chiral ligand-exchange chromatography for the enantioseparation of propranolol
A new chromatographic procedure was proposed for the separation of propranolol (PRN) enantiomers based upon enantioselective chiral ligand-exchange chromatography. The separation was carried out on a short C8 column leading to considerably short separation time. L-alanine and Cu2+ were applied as chiral selector and central bivalent complexing ion, respectively. It was found that the kind of co...
متن کاملDevelopment of a new method based on chiral ligand-exchange chromatography for the enantioseparation of propranolol
A new chromatographic procedure was proposed for the separation of propranolol (PRN) enantiomers based upon enantioselective chiral ligand-exchange chromatography. The separation was carried out on a short C8 column leading to considerably short separation time. L-alanine and Cu2+ were applied as chiral selector and central bivalent complexing ion, respectively. It was found that the kind of co...
متن کاملReview High-performance liquid chromatographic enantioseparation of drugs containing multiple chiral centers on chiral stationary phases
In recent years there has been considerable interest in the synthesis and separation of enantiomers of organic compounds especially because of their importance in the biochemistry and pharmaceutical industry. High-performance liquid Chromatography is a very useful method for the direct separation of enantiomers. However, about 30−40 years ago, commercially available chiral stationary phases wer...
متن کاملSeparation of Somatropin and Its Degradation Products by High-Performance Liquid Chromatography Using a Reversed-Phase Polymeric Column
The accurate prediction of protein stability is one of the most challenging goals in protein formulation and delivery. In this study, a gradient RP-HPLC method is described for the separation of human growth hormone (hGH) variants as deamidated and oxidized forms. The methodology employed a polymeric poly (styrene-co-divinylbenzene) column and a 1mL/min flow rate of a linear gradient of 0.1% v/...
متن کاملSeparation of Somatropin and Its Degradation Products by High-Performance Liquid Chromatography Using a Reversed-Phase Polymeric Column
The accurate prediction of protein stability is one of the most challenging goals in protein formulation and delivery. In this study, a gradient RP-HPLC method is described for the separation of human growth hormone (hGH) variants as deamidated and oxidized forms. The methodology employed a polymeric poly (styrene-co-divinylbenzene) column and a 1mL/min flow rate of a linear gradient of 0.1% v/...