Evaluation of the effect of gain on the meal response of an automated closed-loop insulin delivery system.
نویسندگان
چکیده
A continuous closed-loop insulin delivery system using subcutaneous insulin delivery was evaluated in eight diabetic canines. Continuous glucose profiles were obtained by extrapolation of blood glucose measurements. Insulin delivery rate was calculated, using a model of beta-cell insulin secretion, and delivered with a Medtronic MiniMed subcutaneous infusion pump. The model acts like a classic proportional-integral-derivative controller, delivering insulin in proportion to glucose above target, history of past glucose values, and glucose rate of change. For each dog, a proportional gain was set relative to the open-loop total daily dose (TDD) of insulin. Additional gains based on 0.5 x TDD and 1.5 x TDD were also evaluated (gain dose response). Control was initiated 4 h before the meal with a target of 6.7 mmol/l. At the time of the meal, glucose was similar for all three gains (6.0 +/- 0.3, 5.2 +/- 0.3, and 4.9 +/- 0.5 mmol/l for 0.5 x TDD, TDD, and 1.5 x TDD, respectively; P > 0.05) with near-target values restored at the end of experiments (8.2 +/- 0.9, 6.0 +/- 0.6, and 6.0 +/- 0.5, respectively). The peak postprandial glucose level decreased significantly with increasing gain (12.1 +/- 0.6, 9.6 +/- 1.0, and 8.5 +/- 0.6 mmol/l, respectively; P < 0.05). The data demonstrate that closed-loop insulin delivery using the subcutaneous site can provide stable glycemic control within a range of gain.
منابع مشابه
Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes
OBJECTIVE To minimize hypoglycemia in subjects with type 1 diabetes by automated glucagon delivery in a closed-loop insulin delivery system. RESEARCH DESIGN AND METHODS Adult subjects with type 1 diabetes underwent one closed-loop study with insulin plus placebo and one study with insulin plus glucagon, given at times of impending hypoglycemia. Seven subjects received glucagon using high-gain...
متن کاملControl of Flexible Link Robot using a Closed Loop Input-Shaping Approach
This paper is has addressed the Single Flexible Link Robot. The dynamical model is derived using Euler-Lagrange equation and then a proper controller is designed to suppress a vibration based-on Input-Shaping (IS) method. But, IS control method is an open loop strategy. Due to the weakness of open loop control systems, a closed loop IS control system is proposed. The achieved closed loop c...
متن کاملA New Near Optimal High Gain Controller For The Non-Minimum Phase Affine Nonlinear Systems
In this paper, a new analytical method to find a near-optimal high gain controller for the non-minimum phase affine nonlinear systems is introduced. This controller is derived based on the closed form solution of the Hamilton-Jacobi-Bellman (HJB) equation associated with the cheap control problem. This methodology employs an algebraic equation with parametric coefficients for the systems with s...
متن کاملA New Approach in Preliminary Design of Closed Loop Solar Thermal Systems (RESEARCH NOTE)
In this paper, a model for closed loop solar system is presented and an attempt is made to generalize the model to be utilized for primary design of any solar active thermal system. This model may be used for systems in which gas or a liquid are fluids that flow. Two new parameters, namely, the system heat delivery factor and the system heat absorption factor are introduced in the model. These ...
متن کاملThe artificial pancreas: is it important to understand how the β cell controls blood glucose?
It has been more than 7 years since the first fully automated closed-loop insulin delivery system that linked subcutaneous insulin delivery and glucose sensing was published. Since the initial report, the physiologic insulin delivery (PID) algorithm used to emulate the β cell has been modified from the original proportional-integral-derivative terms needed to fit the β cell's biphasic response ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Diabetes
دوره 55 7 شماره
صفحات -