High resolution gel-dosimetry by optical-CT and MR scanning.

نویسندگان

  • M Oldham
  • J H Siewerdsen
  • A Shetty
  • D A Jaffray
چکیده

The increased intricacy of Intensity-Modulated-Radiation-Therapy (IMRT) delivery has created the need for a high-resolution 3D-dosimetry (three-dimensional) system capable of measuring and verifying the complex delivery. Present clinical methods are inadequate being restricted to single points (e.g., ion-chambers) or to 2D planes (e.g., film), and are labor intensive. In this paper we show that gel-dosimetry in conjunction with optical-CT scanning can yield maps of dose that are of sufficient accuracy, resolution and precision to allow verification of complex radiosurgery deliveries, and by extension IMRT deliveries. The radiosurgery dose-distribution represents the most challenging case encountered in external beam therapy by virtue of the steep dose-gradients and high resolution of delivery. We characterize the stringent radiosurgery requirements by the RTAP (Resolution-Time-Accuracy-Precision) criteria defined as < or = 1 mm3 spatial resolution, < or = 1 hour imaging time, accurate to within 3%, and within -1% precision. The RTAP criteria is applied to an in-house laser-based optical-CT scanning system presented here, and evaluated using gel-flasks containing BANG3 gel. The same gel flasks were subsequently imaged using the MR imaging protocol recommended by the gel manufacturer, but modified to match as closely as possible the RTAP. The resulting dose-maps demonstrate the high precision (< 1.3% noise at high dose) achievable with optical CT scanning while preserving high spatial resolution (<1 mm3). Using the sequence above, the MR gel-dose maps were found to have poorer precision by a factor of 5, under the strict conditions of the RTAP. The optical CT gel-dosimetry system was further evaluated for the verification of a complex 3-isocenter radiosurgery delivery. In conclusion, this work demonstrates that gel-dosimetry and optical-CT scanning approach an important long-term goal of radiation dosimetry, as specified by the RTAP criteria, and have potential to impact the clinic by improving and facilitating clinical dose verification for the most complex external beam radiation treatments.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Investigation of the Spatial Resolution of MR-Based Polymer Gel Dosimetry versus Film Densitometry using Dose Modulation Transfer Function

Introduction: The conventional methods of dosimetry are not capable of dosimetry in such a small volume of less than one cubic millimeter. Although the polymer gel dosimetry method based on magnetic resonance imaging (MRI) could achieve three dimensional dosimetry with high resolution, a spatial resolution evaluation based on gel dose modulation transfer function has not been investigated yet. ...

متن کامل

Designing and developing an in-house CCD based optical CT scanner for gel dosimetry

Introduction: Measurement of the complex 3D radiation dose distribution created with advanced radiotherapy techniques such as IMRT and SRS became very important. Polymer gel dosimetry (PGD) is proposed method to recording this dose distribution. MRI, X-ray CT, ultrasound, spectrophotometry and optical CT scanners can be used for readout of Polymer gel dosimeters. Although MRI i...

متن کامل

Uncertainty in 3D gel dosimetry

Three-dimensional (3D) gel dosimetry has a unique role to play in safeguarding conformal radiotherapy treatments as the technique can cover the full treatment chain and provides the radiation oncologist with the integrated dose distribution in 3D. It can also be applied to benchmark new treatment strategies such as image guided and tracking radiotherapy techniques. A major obstacle that has hin...

متن کامل

Fan-beam scanning laser optical computed tomography for large volume dosimetry

A prototype scanning-laser fan beam optical CT scanner is reported which is capable of high resolution, large volume dosimetry with reasonable scan time. An acylindrical, asymmetric aquarium design is presented which serves to 1) generate parallel-beam scan geometry, 2) focus light towards a small acceptance angle detector, and 3) avoid interference fringe-related artifacts. Preliminary experim...

متن کامل

Evaluation of Ferrous-Agarose-Xylenol Gel Properties in Radiation Dosimetry

Background: Over recent decades, modern protocols of external beam radioÂ‌therapy and radiation techniques such as intensity-modulated radiotherapy (IMRT) have been developed. These methods are extremely sensitive to errors in treatment delivery, so that it is essential to apply a high resolution 3D dosimetry system that has high sensitivity and is capable of measuring and verifying the complex...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 28 7  شماره 

صفحات  -

تاریخ انتشار 2001