Analytical Compensation for Spatially Variant Detector Response in SPECT With Varying Focal-Length Fan-Beam Collimators
نویسندگان
چکیده
The authors present a new method to compensate for the spatially variant detector response in single photon emission computed tomography (SPECT) with varying focal-length fan-beam (VFF) geometry. In this method, the deblurred projection is explicitly expressed as an integral of the collimator blurred data and the fully calibrated system kernel matrix with certain transforms. It needs a little more computation time than the conventional filtering techniques but provides much more accurate compensation. This method also represents a unified framework for detector response compensation in parallel-hole fan-beam and VFF SPECT with a circular scan orbit.
منابع مشابه
Analytical Solution to 3D SPECT Reconstruction with Non-uniform Attenuation, Scatter, and Spatially-variant Resolution Variation for Variable Focal-length Fan-beam Collimators
In the past decades, analytical (non-iterative) methods have been extensively investigated and developed for the reconstruction of three-dimensional (3D) single-photon emission computed tomography (SPECT). However, it becomes possible only recently when the exact analytic non-uniform attenuation reconstruction algorithm was derived. Based on the explicit inversion formula for the attenuated Rad...
متن کاملAn Analytical Inversion of the Nonuniformly Attenuated Radon Transform With Variable Focal-Length Fan-Beam Collimators
Single photon emission computed tomography (SPECT) is based on the measurement of radiation emitted by a radiotracer injected into the patient. Because of photoelectric absorption and Compton scatter, the gamma photons are attenuated inside the body before arriving at the detector. A quantitative reconstruction must consider the attenuation, which is usually nonuniform. Novikov has derived an e...
متن کاملFan-beam reconstruction algorithm for a spatially varying focal length collimator
Fan-beam collimators are used in single-photon-emission computed tomography (SPECT) to improve the sensitivity for imaging of small organs. The disadvantage of fan-beam collimation is the truncation of projection data surrounding the organ of interest or, in those cases of imaging large patients, of the organ itself, producing reconstruction artifacts. A spatially varying focal length fan-beam ...
متن کاملRay-driven analytical fan-beam SPECT reconstruction with nonuniform attenuation
Single-photon emission computed tomography (SPECT) is based on the measurement of radiation emitted by a radiotracer injected into the patient. Because of photoelectric absorption and Compton scatter, the γ photons are attenuated inside the body before arriving at the detector. A quantitative reconstruction must consider the attenuation, which is usually non-uniform. Novikov and Natterer had de...
متن کاملAn inversion formula for the exponential radon transform in spatial domain with variable focal-length fan-beam collimation geometry.
Inverting the exponential Radon transform has a potential use for SPECT (single photon emission computed tomography) imaging in cases where a uniform attenuation can be approximated, such as in brain and abdominal imaging. Tretiak and Metz derived in the frequency domain an explicit inversion formula for the exponential Radon transform in two dimensions for parallel-beam collimator geometry. Pr...