Comparison of force attenuation properties of four different hip protectors under simulated falling conditions in the elderly: an in vitro biomechanical study.

نویسندگان

  • P Kannus
  • J Parkkari
  • J Poutala
چکیده

The purpose of this in vitro biomechanical study is to determine the force attenuation capacity of four different hip protectors (KPH1, KPH2, Safehip, and Safetypants) in falling simulations in elderly subjects (falls to the side). The simulated falling conditions were created by a biomechanical testing system, which consisted of an impact pendulum, surrogate pelvis and femur, and two load cells. Three series of impact experiments were conducted in an ascending order (low, moderate-, and high-force experiments), each exceeding the literature-provided average (+/- 1 SD) fracture threshold (3100 +/- 1200 N) of the proximal femur of elderly women with a mean age of 71 years. Using a low impact force of 4330 N, the trochanteric soft tissue (20-mm-thick polyethylene foam) attenuated the peak femoral impact force to 3740 N and, accordingly, the KPH1 protector to 590 N, KPH2 to 510 N, Safehip to 1080 N, and Safetypants to 790 N. Thus, in this low force experiment, all tested protectors could reduce the peak impact force entered into the proximal femur below the aforementioned average fracture threshold area (3100 +/-1200 N) of the proximal femur of elderly women. With a moderate impact force of 7230 N, the soft tissue attenuated the peak femoral impact force to 6130 N, and the protectors to 780 N, 760 N, 2240 N, and 2760 N, respectively. Thus, with this impact force, only the KPH hip protectors could reduce the impact force clearly below the fracture threshold area. In the final series of the experiment, the peak femoral impact force was set to be so high (10,840 N) that the protector, if effective, should prevent the hip fracture in almost all cases and situations. The trochanteric soft tissue attenuated this peak impact force to 9190 N, and the tested protectors to 1360 N, 1170 N, 4640 N, and 5770 N. Thus, with the KPH protectors the force received by the proximal femur remained below the average force required to fracture the proximal femur of elderly women, whereas with the two other protectors the impact force entered into the proximal femur clearly exceeded this threshold value. In conclusion, the test results showed that, of the four tested hip protectors, the anatomically designed energy-shunting and energy-absorbing KPH protectors can provide an effective impact force attenuation in a sideways-fall simulation in the elderly, whereas the force attenuation capacity of the two other protectors seems more limited. However, the true efficacy of any protector in the prevention of hip fractures can only be evaluated in randomized clinical trials.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical comparison of hard and soft hip protectors, and the influence of soft tissue.

INTRODUCTION Hip protectors appear to be promising in preventing hip fractures. Currently, many different hip protectors exist, and it is not clear which hip protector has the best biomechanical properties. Therefore, the objective of this study was to compare the force attenuation capacity of 10 different hip protectors. Both hard hip protectors, which primarily shunt away energy, and soft hip...

متن کامل

The effects of pad geometry and material properties on the biomechanical effectiveness of 26 commercially available hip protectors.

Wearable hip protectors (padded garments) represent a promising strategy to decrease impact force and hip fracture risk during falls, and a wide range of products are currently marketed. However, little is known about how design features of hip protectors influence biomechanical effectiveness. We used a mechanical test system (simulating sideways falls) to measure the attenuation in femoral nec...

متن کامل

Biomechanical study of an anthropometrically designed hip protector for older chinese women.

Biomechanical properties and compliance are 2 essential properties of hip protectors (HPs) that determine their clinical efficacy. This study describes the development of a HP that was anthropometrically designed according to anthropometric data from 68 older Chinese women and with attention paid to the biomechanical properties of the device. A simulated mechanical fall test using a drop weight...

متن کامل

The Comparison of Falling Risk of Elderly by Speed Gait Test Under Dual Tasks Conditions

Objectives: This study aimed to compare elderly fallers and non fallers by balance test under dual tasks conditions. Methods & Materials: This study was a analyse-comparative study. Subjects were chosen from three parks of Tehran. Subjects were 20 older adults with no history of falls (aged 72.60±5 years) and 21 older adults with a history of 2 or more falls in the last one year (aged 74.50±...

متن کامل

The Relationship Between Muscle Strength and Range of Motion in Lower Extremity, and Balance and Risk of Falling in the Elderly

Background: The present study examined the relationship between muscle strength and range of motion in the lower extremity, and balance and risk of falling in the elderly. Methods: This cross-sectional study was conducted on healthy elderly (mean±SD age: 65.64±4.95 y) in Rasht City, Iran. The study participants were 65 healthy male elderly who met the stud inclusion criteria. They were selecte...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Bone

دوره 25 2  شماره 

صفحات  -

تاریخ انتشار 1999