Inspiratory limb carbon dioxide entrainment during high-frequency oscillatory ventilation: characterization in a mechanical test lung and swine model.

نویسندگان

  • Adam W Bostick
  • Gregory A Naworol
  • Tyler J Britton
  • Timothy R Ori
  • Shawn K French
  • Stephen Derdak
چکیده

BACKGROUND High-frequency oscillatory ventilation (HFOV) has been utilized as a rescue oxygenation therapy in adults with ARDS over the last decade. The HFOV oscillating piston can generate negative pressure during the exhalation cycle, which has been termed active exhalation. We hypothesized that this characteristic of HFOV entrains CO(2) into the inspiratory limb of the circuit and increases the total dead space. The purpose of this study was to determine if retrograde CO(2) entrainment occurs and how it is altered by HFOV parameter settings. METHODS An HFOV was interfaced to a cuffed endotracheal tube and connected to a mechanical test lung. Negative pressure changes within the circuit's inspiratory limb were measured while HFOV settings were manipulated. Retrograde CO(2) entrainment was evaluated by insufflating CO(2) into the test lung to achieve 40 mm Hg at the carina. Inspiratory limb CO(2) entrainment was measured at incremental distances from the Y-piece. HFOV settings and cuff leak were varied to assess their effect on CO(2) entrainment. Control experiments were conducted using a conventional ventilator. Test lung results were validated on a large hypercapnic swine. RESULTS Negative pressure was detectable within the inspiratory limb of the HFOV circuit and varied inversely with mean airway pressure (P(-)(aw)) and directly with oscillatory pressure amplitude (ΔP). CO(2) was readily detectable within the inspiratory limb and was proportional to the negative pressure that was generated. Factors that decreased CO(2) entrainment in both the test lung and swine included low ΔP, high mean airway pressure, high oscillatory frequency (Hz), high bias flow, and endotracheal tube cuff leak placement. CO(2) entrainment was also reduced by utilizing a higher bias flow strategy at any targeted mean airway pressure. CONCLUSIONS Retrograde CO(2) entrainment occurs during HFOV use and can be manipulated with the ventilator settings. This phenomenon may have clinical implications on the development or persistence of hypercapnia.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Carbon dioxide entrainment during high-frequency oscillatory ventilation: a model analysis.

High-frequency ventilation got its start in the United States from a paper by Dr Mirosolav Klain, an anesthesiologist who described a high-frequency jet ventilator (HFJV)1 he built using a fluidic control circuit purchased as a demonstration kit from Corning Glass Works.2 Based on his description of the circuit, we built an HFJV and humidification system that we used to ventilate neonates clini...

متن کامل

Clinical experience with a pumpless extracorporeal lung assist device.

We present three patients with respiratory failure in whom conventional mechanical lung ventilation resulted in unacceptably high levels of carbon dioxide, severe acidosis and high vasopressor requirements. A pumpless arteriovenous extracorporeal carbon dioxide removal device (Novalung) was inserted. Arterial carbon dioxide levels were reduced rapidly with a corresponding increase in pH, reduct...

متن کامل

Jet Ventilation during Rigid Bronchoscopy in Adults: A Focused Review

The indications for rigid bronchoscopy for interventional pulmonology have increased and include stent placements and transbronchial cryobiopsy procedures. The shared airway between anesthesiologist and pulmonologist and the open airway system, requiring specific ventilation techniques such as jet ventilation, need a good understanding of the procedure to reduce potentially harmful complication...

متن کامل

Moderately high frequency ventilation with a conventional ventilator allows reduction of tidal volume without increasing mean airway pressure

BACKGROUND The aim of this study was to explore if positive-pressure ventilation delivered by a conventional ICU ventilator at a moderately high frequency (HFPPV) allows a safe reduction of tidal volume (V T) below 6 mL/kg in a porcine model of severe acute respiratory distress syndrome (ARDS) and at a lower mean airway pressure than high-frequency oscillatory ventilation (HFOV). METHODS This...

متن کامل

Aerosol delivery during high frequency jet ventilation: an MRI evaluation.

BACKGROUND We have previously demonstrated aerosol delivery during conventional and high frequency oscillatory (HFOV) ventilation using magnetic resonance imaging (MRI) in piglets. There are no reports on aerosol delivery during high frequency jet ventilation (HFJV). OBJECTIVE To compare delivery of aerosolized gadopentetate dimeglumine (Gd-DTPA) in 3 neonatal ventilator circuits: convention...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Respiratory care

دوره 57 11  شماره 

صفحات  -

تاریخ انتشار 2012