Non-Contact Measurement of the Spectral Emissivity through Active/Passive Synergy of CO2 Laser at 10.6 µm and 102F FTIR (Fourier Transform Infrared) Spectrometer
نویسندگان
چکیده
In the inversion of land surface temperature (LST) from satellite data, obtaining the information on land surface emissivity is most challenging. How to solve both the emissivity and the LST from the underdetermined equations for thermal infrared radiation is a hot research topic related to quantitative thermal infrared remote sensing. The academic research and practical applications based on the temperature-emissivity retrieval algorithms show that directly measuring the emissivity of objects at a fixed thermal infrared waveband is an important way to close the underdetermined equations for thermal infrared radiation. Based on the prior research results of both the authors and others, this paper proposes a new approach of obtaining the spectral emissivity of the object at 8-14 µm with a single-band CO₂ laser at 10.6 µm and a 102F FTIR spectrometer. Through experiments, the spectral emissivity of several key samples, including aluminum plate, iron plate, copper plate, marble plate, rubber sheet, and paper board, at 8-14 µm is obtained, and the measured data are basically consistent with the hemispherical emissivity measurement by a Nicolet iS10 FTIR spectrometer for the same objects. For the rough surface of materials, such as marble and rusty iron, the RMSE of emissivity is below 0.05. The differences in the field of view angle and in the measuring direction between the Nicolet FTIR method and the method proposed in the paper, and the heterogeneity in the degree of oxidation, polishing and composition of the samples, are the main reasons for the differences of the emissivities between the two methods.
منابع مشابه
Remote sensing of atmospheric pollution by passive FTIR spectrometry
Passive remote sensing with a Fourier transform infrared (FTIR) spectrometer allows the detection and identification of pollutant clouds in the atmosphere. In this work, the measurement technique and a data analysis method that does not require a previously measured background spectrum are described. Recent experimental results obtained with a new high sensitive FTIR remote sensor are presented...
متن کاملMeasurement of Combustion Temperature of Solid Propellant by FTIR Spectrometer
The temperature of solid state propellant combustion flame is studied by a Bruker OPAG33 Fourier Transform Infrared Spectrometer (FTIR). The infrared emission spectra of Nitrate Ester Plasticized Polyether Propellant (NEPE) solid propellant are measured with a resolution of 4 cm−1 at a testing distance of 20 m in real time. The temperature of NEPE propellant plume is calculated by using the fin...
متن کاملFourier-Transform Infrared Spectroscopic Comparison of Normal and Malignant Cervical Tissue Cervical Tissue.
Based on the accumulated data, FTIR spectroscopy seems to be a highly sensitive tool in cervical cancer diagnosis. We, therefore, employed this technique in a screening type of research here in Tehran. The purpose of this study was to look for the spectral pattern differences between normal and malignant cervix samples in Iranian women. Through formal and informal announcements, all gynecology ...
متن کاملComparison of field and laboratory collected midwave and longwave infrared emissivity spectra / data reduction techniques
Many targets that remote sensing scientists encounter when conducting their research experiments do not lend themselves to laboratory measurement of their surface optical properties. Removal of these targets from the field can change their biotic condition, disturb the surface composition, and change the moisture content of the sample. These parameters, as well as numerous others, have a marked...
متن کاملNovel high-temperature and pressure-compatible ultrasonic levitator apparatus coupled to Raman and Fourier transform infrared spectrometers.
We describe an original apparatus comprising of an acoustic levitator enclosed within a pressure-compatible process chamber. To characterize any chemical and physical modifications of the levitated particle, the chamber is interfaced to complimentary, high-sensitivity Raman (4390-170 cm(-1)), and Fourier transform infrared (FTIR) (10,000-500 cm(-1)) spectroscopic probes. The temperature of the ...
متن کامل