Determination of parameters for a multiple-source model of megavoltage photon beams using optimization methods.

نویسندگان

  • L Tillikainen
  • S Siljamäki
  • H Helminen
  • J Alakuijala
  • J Pyyry
چکیده

Accurate modelling of the radiation output of a medical linear accelerator is important for radiotherapy treatment planning. The major challenge is the adjustment of the model to a specific treatment unit. One approach is to use a multiple-source model containing a set of physical parameters. In this work, the parameters were derived from standard beam data measurements using optimization methods. The source model used includes sub-sources for bremsstrahlung radiation from the target, extra-focal photon radiation and electron contamination. The cost function includes a gamma error measure between measurements and current dose calculations. The procedure was applied to six beam data sets (6 MV to 23 MV) measured with accelerators from three vendors, but the results focus primarily on Varian accelerators. The obtained average gamma error (1%, 1 mm) between dose calculations and measurements used in optimization was smaller than 0.7 for each studied treatment beam and field size, and a minimum of 83% of measurement points passed the gamma < 1 criterion. For experiments made at different SSDs and for asymmetric fields, the average gamma errors were smaller than 1.1. For irregularly shaped MLC apertures, the differences in point doses were smaller than 1.0%. This work demonstrates that the source model parameters can be automatically derived from simple measurements using optimization methods. The developed procedure is applicable to a wide range of accelerators, and has an acceptable accuracy and processing time.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calculating Weighting Factors for Mixing Megavoltage Photon Beams to Achieve Desirable Dose Distribution in Radiotherapy

Background: In radiotherapy, low-energy photon beams are better adapted to the treated volume, and the use of high-energy beams can reduce hot spots in the radiation therapy. Therefore, mixing low and high energies with different ratios can control the rate of hotspots, as well as the dose distribution of the target volume.Material and Methods: The percentage depth doses (PDDs) were calculated ...

متن کامل

Evaluation of Electron Contamination in Cancer Treatment with Megavoltage Photon Beams: Monte Carlo Study

Background: Megavoltage beams used in radiotherapy are contaminated with secondary electrons. Different parts of linac head and air above patient act as a source of this contamination. This contamination can increase damage to skin and subcutaneous tissue during radiotherapy. Monte Carlo simulation is an accurate method for dose calculation in medical dosimetry and has an important role in opt...

متن کامل

Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Introduction For dose measurement in Megavoltage (MV) photon beams with ion chambers, the effect of volume occupied by the air cavity is not negligible. Therefore, the result of measurement should be corrected with a displacement perturbation correction factor (Pdis) or using an effective point of measurement (EPOM). The aim of this study is to calculate the EPOM for cylindrical ion chamber and...

متن کامل

Dosimetric comparison between XR-RV3 and EBT2 radiochromic film in megavoltage photon beams

Background:  Radiochromic film is used for radiation dose measurement, XR-RV3 is used in fluoroscopy and EBT2 film in radiation therapy. The aim was to determine if the dosimetric properties of these two films are comparable in megavolt photon beams. Materials and Methods: Comparison measurements included: calibration curves, heterogeneous phantom dose profiles, and nasopharynx dose distributio...

متن کامل

Determination of an Effective Wedge Angle by Combination of Two Arbitrary Universal Wedge Fields in Radiation Therapy of Cancer Patients with Megavoltage Photon Beams.

Introduction: Wedge filters are commonly used in radiation oncology for eliminating hot spots and creating a uniform dose distribution in optimizing isodose curves in the target volume for clinical aspects. These are some limited standard physical wedges (15°, 30°, 45°, 60°), or creating an arbitrary wedge angle, like motorized wedge or dynamic wedge, ... The new formulation is...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 52 5  شماره 

صفحات  -

تاریخ انتشار 2007