The Sysmex XN-2000 Hematology Autoanalyzer Provides a Highly Accurate Platelet Count than the Former Sysmex XE-2100 System Based on Comparison with the CD41/CD61 Immunoplatelet Reference Method of Flow Cytometry
نویسندگان
چکیده
Dear Editor The recently launched hematology autoanalyzer Sysmex XN2000 (Sysmex, Kobe, Japan) provides a new method (PLT-F method) for platelet counting based on a fluorescent RNA staining dye named oxazine [1, 2]. The main advantage of the XN2000’s PLT-F method over the former instrument XE-2100’s optical method (PLT-O method) is the ability to identify apoptotic white cells and not count them as platelets [2]. The XN-2000’s PLT-F method was reported to estimate platelets in thrombocytopenic samples more accurately than does the PLT-O method [1]. However, the performance of XN-2000’s PLT-F method still needs to be evaluated, including samples with broad range of platelet counts. We performed correlation and bias analyses comparing the platelet counts measured by the XN-2000 and XE-2100 autoanalyzers with those calculated by the immunoplatelet reference method (IRM) of flow cytometry. In total, 195 peripheral blood samples, which had been submitted for complete blood counts (CBC) at the authors’ institution between July and September 2013, were randomly selected for this study. All samples were analyzed using the XE2100 (PLT-O method) and XN-2000 (PLT-F method) autoanalyzers within 4 hr of sample collection. In addition, the platelet count was acquired by measuring CD41/CD61 IRM of flow cytometry using a FACSCanto II (Becton-Dickinson, Sunnyvale, CA, USA) [3]. This method uses the red blood cell (RBC): platelet ratio, which is calculated from RBC and platelet events measured by flow cytometry (Fig. 1A). Linear regression and Bland– Altman analysis were applied to compare the platelet counts measured by the two hematology autoanalyzers and those produced by CD41/CD61 IRM of flow cytometry. The platelet count-related parameters of the 195 samples obtained from the hematology autoanalyzers and by flow cytometry
منابع مشابه
Platelet counting by the coulter LH 750, sysmex XE 2100, and advia 120: a comparative analysis using the RBC/platelet ratio reference method.
We compared the accuracy and precision of the impedance platelet counts generated by the Beckman Coulter LH 750 and the Sysmex XE 2100 and the optical platelet counts produced by the Advia 120 and the Sysmex XE 2100 withflow cytometric reference platelet counts. Samples analyzed had platelet counts less than 150 x 10(3)/microL (150 x 10(9)/L) with a platelet flag or less than 75 x 10(3)/microL ...
متن کاملThe New Sysmex XN-2000 Automated Blood Cell Analyzer More Accurately Measures the Absolute Number and the Proportion of Hematopoietic Stem and Progenitor Cells Than XE-2100 When Compared to Flow Cytometric Enumeration of CD34+ Cells
The gold standard to count circulating hematopoietic stem and progenitor cells (HPCs) is to enumerate CD34 cells by using flow cytometry [1, 2]. XE-2100 (Sysmex, Kobe, Japan), a currently used hematology analyzer, can be used to quantify HPCs, generating counts that reportedly correlate with those obtained using flow cytometric analysis of CD34 cells; however, XE-2100 HPC counts were 1.86-fold ...
متن کاملNew quantitative parameters on a recently introduced automated blood cell counter--the XE 2100.
The XE 2100 (Sysmex Corporation) is a cell counter that furthers the technology of fluorescent flow cytometry developed from the earlier range of Sysmex analysers. The new diagnostic features are a nucleated red cell count (NRBC), the ability to measure platelets by impedance as well as an 'optical' platelet count using a fluorescence dye and an immature granulocyte (IG) count. The NRBC count w...
متن کاملAccuracy of a New Platelet Count System (PLT-F) Depends on the Staining Property of Its Reagents
BACKGROUND Platelet count is essential for the diagnosis and management of hemostasis abnormalities. Although existing platelet count methods installed in common hematology analyzers can correctly count platelets in normal blood samples, they tend to miscount platelets in some abnormal samples. The newly developed PLT-F channel in the XN-Series hematology analyzer (Sysmex) has been reported to ...
متن کاملEvaluation of the Diagnostic Performance of theSysmex XT-2000i Automated Hematology Analyzerin the Detection of Immature Granulocyte
Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microsco...