Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present.
نویسندگان
چکیده
BACKGROUND Although evaluation of cardiac output by the partial carbon dioxide rebreathing technique is as accurate as thermodilution techniques under controlled mechanical ventilation, it is less accurate at low tidal volume. It is not clear whether reduced accuracy is due to low tidal volume or low minute ventilation. The effect of spontaneous breathing on the accuracy of partial carbon dioxide rebreathing measurement has not been fully investigated. The objectives of the current study were to investigate whether tidal volume or minute ventilation is the dominant factor for the accuracy, and the accuracy of the technique when spontaneous breathing effort is present. METHODS The authors enrolled 25 post-cardiac surgery patients in two serial protocols. First, the authors applied three settings of controlled mechanical ventilation in random order: large tidal volume (12 ml/kg), the same minute ventilation with a small tidal volume (6 ml/kg), and 50% decreased minute ventilation with a small tidal volume (6 ml/kg). Second, when the patient recovered spontaneous breathing, the authors applied three conditions of partial ventilatory support in random order: synchronized intermittent mandatory ventilation-pressure support ventilation, pressure support ventilation with an appropriately adjusted rebreathing loop, and pressure support ventilation with the shortest available loop. After establishing steady state conditions, the authors measured cardiac output using both partial carbon dioxide rebreathing and thermodilution methods. The correlation between the data yielded by the two methods was determined by Bland-Altman analysis and linear regression. RESULTS Cardiac output with the carbon dioxide rebreathing technique correlated moderately with that measured by thermodilution when minute ventilation was set to maintain normocapnia, regardless of tidal volumes. However, when minute ventilation was set low, the carbon dioxide rebreathing technique underreported cardiac output (y = 0.70x; correlation coefficient, 0.34; bias, -1.73 l/min; precision, 1.27 l/min; limits of agreement, -4.27 to +0.81 l/min). When there was spontaneous breathing, the correlation between the two cardiac output measurements became worse. Carbon dioxide rebreathing increased spontaneous tidal volume and respiratory rate (20% and 30%, respectively, during pressure support ventilation) when the rebreathing loop was adjusted for large tidal volume. CONCLUSIONS During controlled mechanical ventilation, minute ventilation rather than tidal volume affected the accuracy of cardiac output measurement using the partial carbon dioxide rebreathing technique. When spontaneous breathing is present, the carbon dioxide rebreathing technique is less accurate and increases spontaneous tidal volume and respiratory rate.
منابع مشابه
Effects of reduced rebreathing time, in spontaneously breathing patients, on respiratory effort and accuracy in cardiac output measurement when using a partial carbon dioxide rebreathing technique: a prospective observational study
INTRODUCTION New technology using partial carbon dioxide rebreathing has been developed to measure cardiac output. Because rebreathing increases respiratory effort, we investigated whether a newly developed system with 35 s rebreathing causes a lesser increase in respiratory effort under partial ventilatory support than does the conventional system with 50 s rebreathing. We also investigated wh...
متن کاملEffect of ventilatory settings on accuracy of cardiac output measurement using partial CO(2) rebreathing.
BACKGROUND Recently, a new device has been developed to measure cardiac output noninvasively using partial carbon dioxide (CO(2)) rebreathing. Because this technique uses CO(2) rebreathing, the authors suspected that ventilatory settings, such as tidal volume and ventilatory mode, would affect its accuracy: they conducted this study to investigate which parameters affect the accuracy of the mea...
متن کاملSources of error in noninvasive pulmonary blood flow measurements by partial rebreathing: a computer model study.
BACKGROUND Partial rebreathing is a noninvasive method for measuring pulmonary blood flow (PBF). This study examines the systematic errors produced by the partial rebreathing technique utilizing a comprehensive mathematical model of the cardiorespiratory system of a healthy, 70-kg adult male. METHODS The model simulates tidal breathing through a branched respiratory tree and incorporates the ...
متن کاملComparison between cardiac output values measured by thermodilution and partial carbon dioxide rebreathing in patients with acute lung injury.
CONTEXT Thermodilution, which is considered to be a standard technique for measuring the cardiac output in critically ill patients, is not free from relevant risks. There is a need to find alternative, noninvasive, automatic, simple and accurate methods for monitoring cardiac output at the bedside. OBJECTIVE To compare cardiac output measurements by thermodilution and partial carbon dioxide r...
متن کاملNoninvasive monitoring cardiac output using partial CO(2) rebreathing.
This article reviews use of partial carbon dioxide rebreathing devices to determine cardiac output and their application for hemodynamic monitoring in the ICU and operating room. The primary focus is on the NICO monitoring device. Compared with conventional cardiac output methods, these techniques are noninvasive, easily automated, and provide real-time and continuous cardiac output monitoring....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anesthesiology
دوره 98 4 شماره
صفحات -
تاریخ انتشار 2003