Effects of changing modulation and pitch parameters on tomotherapy delivery quality assurance plans
نویسندگان
چکیده
This study was aimed at investigating delivery quality assurance (DQA) discrepancies observed for helical tomotherapy plans. A selection of tomotherapy plans that initially failed the DQA process was chosen for this investigation. These plans failed the fluence analysis as assessed using gamma criteria (3%, 3 mm) with radiographic film. Each of these plans was modified (keeping the planning constraints the same), beamlets rebatched and reoptimized. By increasing and decreasing the modulation factor, the fluence in a circumferential plane as measured with a diode array was assessed. A subset of these plans was investigated using varied pitch values. Metrics for each plan that were examined were point doses, fluences, leaf opening times, planned leaf sinograms, and uniformity indices. In order to ensure that the treatment constraints remained the same, the dose-volume histograms (DVHs) of all the modulated plans were compared to the original plan. It was observed that a large increase in the modulation factor did not significantly improve DVH uniformity, but reduced the gamma analysis pass rate. This also increased the treatment delivery time by slowing down the gantry rotation speed which then increases the maximum to mean non-zero leaf open time ratio. Increasing and decreasing the pitch value did not substantially change treatment time, but the delivery accuracy was adversely affected. This may be due to many other factors, such as the complexity of the treatment plan and site. Patient sites included in this study were head and neck, right breast, prostate, abdomen, adrenal, and brain. The impact of leaf timing inaccuracies on plans was greater with higher modulation factors. Point-dose measurements were seen to be less susceptible to changes in pitch and modulation factors. The initial modulation factor used by the optimizer, such that the TPS generated 'actual' modulation factor within the range of 1.4 to 2.5, resulted in an improved deliverable plan.
منابع مشابه
Evaluation of the optimal combinations of modulation factor and pitch for Helical TomoTherapy plans made with TomoEdge using Pareto optimal fronts
BACKGROUND Modulation factor (MF) and pitch have an impact on Helical TomoTherapy (HT) plan quality and HT users mostly use vendor-recommended settings. This study analyses the effect of these two parameters on both plan quality and treatment time for plans made with TomoEdge planning software by using the concept of Pareto optimal fronts. METHODS More than 450 plans with different combinatio...
متن کاملEffects of dose scaling on delivery quality assurance in tomotherapy
Delivery quality assurance (DQA) of tomotherapy plans is routinely performed with silver halide film which has a limited range due to the effects of saturation. DQA plans with dose values exceeding this limit require the dose of the entire plan to be scaled downward if film is used, to evaluate the dose distribution in two dimensions. The potential loss of fidelity between scaled and unscaled D...
متن کاملSU-E-T-635: Effect of Planning Parameters on Tomotherapy Dosimetric Quality and Treatment Efficiency.
PURPOSE To investigate how the setting of optimization parameters, fractional dose and tuning structure in tomotherapy treatment planning affects plan dosimetric quality and treatment efficiency. METHODS A digital phantom to simulate head and neck radiotherapy was constructed for this study. Tumor was 10cm long C-shaped with two surrounding parallel normal structures (P-NS) and one serial nor...
متن کاملIs a quasi-3D dosimeter better than a 2D dosimeter for Tomotherapy delivery quality assurance?
Delivery quality assurance (DQA) has been performed for each Tomotherapy patient either using ArcCHECK or MatriXX Evolution in our clinic since 2012. ArcCHECK is a quasi-3D dosimeter whereas MatriXX is a 2D detector. A review of DQA results was performed for all patients in the last three years, a total of 221 DQA plans. These DQA plans came from 215 patients with a variety of treatment sites i...
متن کاملFirst experiences in using a dose control system on a TomoTherapy Hi·Art II
The purpose of this study was to investigate the impact of a dose control system (DCS) servo installed on two fully commissioned TomoTherapy Hi·Art II treatment units. This servo is designed to actively adjust machine parameters to control the output variation of a tomotherapy unit to within ± 0.5% of the nominal dose rate. Machine output, dose rate, and patient-specific quality assurance data ...