Comparison of Aqueous Outflow Facility Measurement by Pneumatonography and Digital Schiøtz Tonography
نویسندگان
چکیده
Purpose It is not known if outflow facilities measured by pneumatonography and Schiøtz tonography are interchangeable. In this study we compared outflow facility measured by pneumatonography to outflow facility measured by digital Schiøtz tonography. Methods Fifty-six eyes from 28 healthy participants, ages 41 to 68 years, were included. Intraocular pressure (IOP) was measured in the sitting and supine positions with a pneumatonometer. With the subject in the supine position, IOP was recorded for 2 minutes by using a pneumatonometer with a 10-g weight and for 4 minutes by using a custom digital Schiøtz tonometer. Outflow facility was determined from the changes in pressure and intraocular volume and a standard assumed ocular rigidity coefficient for each instrument, respectively, and by using an ocular rigidity coefficient calculated by measuring pressure without and with a weight added to the pneumatonometer tip. Results The outflow facility was 0.29 ± 0.09 μL/min/mm Hg by Schiøtz tonography and 0.24 ± 0.08 μL/min/mm Hg by pneumatonography (P < 0.001) when using the standard assumed constant ocular rigidity coefficient. Mean calculated ocular rigidity coefficient was 0.028 ± 0.01 μL-1, and outflow facility determined by using this coefficient was 0.23 ± 0.08 μL/min/mm Hg by Schiøtz tonography and 0.21 ± 0.07 μL/min/mm Hg by pneumatonography (P = 0.003). Outflow facilities measured by the two devices were correlated when the ocular rigidity was assumed (r = 0.60, P < 0.001) or calculated (r = 0.70, P < 0.001). Conclusions Outflow facilities measured by pneumatonography were correlated with those measured by Schiøtz tonography, but Schiøtz tonography reported approximately 10% to 20% higher facilities when using the standard method. When ocular rigidity was determined for each eye, differences were smaller. Measurements from these devices cannot be compared directly.
منابع مشابه
Comparative aqueous outflow facility measurements by pneumatonography and Schiotz tonography.
Tonography was performed on 36 eyes of 15 normal and 3 primary open angle glaucoma patients using pneumatonography and classical Schiotz tonography. The average values of the coefficient of outflow facility (C) for the whole sample were virtually identical with both methods. However, both intersubject and interobserver variability were significantly higher with penumatonography. Although both m...
متن کاملAqueous humor outflow facility by tonography does not change with body position.
PURPOSE Intraocular pressure (IOP) varies with body position and previous research has indicated that most, but not all, of the variation in IOP is due to changes in episcleral venous pressure (EVP). Positional changes in other aqueous humor dynamic parameters may contribute to the change in IOP. The purpose of this study was to investigate the variation of aqueous humor outflow facility with b...
متن کاملComparative human aqueous dynamics study between black and white subjects with glaucoma.
PURPOSE To compare the baseline aqueous humor dynamics in white Caucasians and patients of African origin with previously untreated primary open-angle glaucoma (POAG) or ocular hypertension (OHT). METHODS Ninety-one participants were enrolled in this prospective, observational controlled study: 34 black subjects with POAG or OHT, 32 white Caucasian participants with POAG or OHT, and 12 black ...
متن کاملMechanism of Action of Selective Laser Trabeculoplasty and Predictors of Response
Purpose This study was designed to evaluate the changes in aqueous humor dynamics (AHD) produced by selective laser trabeculoplasty (SLT) and to explore if baseline AHD parameters are predictive of IOP response to SLT. Methods Thirty-one consecutive subjects diagnosed with ocular hypertension or primary open-angle glaucoma scheduled to undergo SLT as their primary IOP-lowering therapy were en...
متن کاملExperimental tonography on enucleated human eyes. I. The validity of Grant's tonography formula.
When newer tables for parameters in Grant's formula for tonography are used, one obtains far higher values for the outflow facility than those generally accepted. The aim of this iinvestigation has, therefore, been to test the validity of the tonography formula. The outflow facility determined by constant-rate perfusion in enucleated eyes has been compared with the results of tonography on the ...