A modified VMAT adaptive radiotherapy for nasopharyngeal cancer patients based on CT-CT image fusion
نویسندگان
چکیده
BACKGROUND To investigate the feasibility and benefits of a modified adaptive radiotherapy (ART) by replanning in the initial CT (iCT) with new contours from a repeat CT (rCT) based on CT-CT image fusion for nasopharyngeal cancer (NPC) patients underwent volumetric modulated arc radiotherapy (VMAT). MATERIALS AND METHODS Nine NPC patients underwent VMAT treatment with a rCT at 23rd fraction were enrolled in this study. Dosimetric differences for replanning VMAT plans in the iCT and in the rCT were compared. Volumetric and dosimetric changes of gross tumor volume (GTV) and organs at risk (OARs) of this modified ART were also investigated. RESULTS No dosimetric differences between replanning in the iCT and in the rCT were observed. The average volume of GTV decreased from 78.83 ± 38.42 cm3 in the iCT to 71.44 ± 37.46 cm3 in the rCT, but with no significant difference (p = 0.42).The average volume of the left and right parotid decreased from 19.91 ± 4.89 cm3 and 21.58 ± 6.16 cm3 in the iCT to 11.80 ± 2.79 cm3 and 13.29 ± 4.17 cm3 in the rCT (both p < 0.01), respectively. The volume of other OARs did not shrink very much. No significant differences on PTVGTV and PTVCTV coverage were observed for replanning with this modified ART. Compared to the initial plans, the average mean dose of the left and right parotid after re-optimization were decreased by 62.5 cGy (p = 0.05) and 67.3 cGy (p = 0.02), respectively, and the V5 (the volume receiving 5 Gy) of the left and right parotids were decreased by 7.8% (p = 0.01) and 11.2% (p = 0.001), respectively. There was no significant difference on the dose delivered to other OARs. CONCLUSION Patients with NPC undergoing VMAT have significant anatomic and dosimetric changes to parotids. Repeat CT as an anatomic changes reference and re-optimization in the iCT based on CT-CT image fusion was accurate enough to identify the volume changes and to ensure safe dose to parotids.
منابع مشابه
Application of computed tomography and magnetic resonance imaging fusion images for delineating gross tumor volume in three-dimensional conformal radiotherapy of nasopharyngeal carcinoma
Background: To evaluate computed tomography (CT) and magnetic resonance imaging (MRI) fusion images for delineating gross tumor volume (GTV) in three-dimensional conformal radiotherapy (3D-CRT) of nasophanrygeal carcinoma (NPC), and compare treatment outcomes between CT- and CT+MRI-based targets. Materials and Methods: A total of 120 NPC patients treated with 3D-CRT were included, in which, 60 ...
متن کاملUsing a photographic image processing method as a part of the QA program in some radiotherapy departments where CTs are involved with COVID-19 patients
Background & Aims: Owing to the significant role of CT images in diagnosis and follow up of patients affected with coronavirus, the imaging section of cancer centers in some countries engaged in providing services to COVID-19 patients. The aim of this study was to introduce a CT-independent photographic-based QA method in some radiotherapy departments where CTs are involved with COVID-19 patien...
متن کاملEvaluation of Volumetric Doses of Organs at Risk in Carcinoma Cervix Patients with HDR Intracavitary Brachytherapy and Comparison of CT-based and Conventional Plans
Background: Brachytherapy treatment planning in cervix carcinoma patients using two dimensional (2D) orthogonal images provides only point dose estimates while CT-based planning provides volumetric dose assessment helping in understanding the correlation between morbidity and the dose to organs at risk (OARs) and treatment volume. Objective: Aim of present study is to compare Internationa...
متن کاملInvestigating the effects of different kernels used for CT image reconstruction on dose distributions in treatment planning of kidney cancer radiotherapy
Introduction: The quality of CT images used for treatment planning of cancer patients is an important issue in accurate outlining of the tumor volume and organs at risk. Different kernels in CT scanner systems are available for improving the image quality. Applying these kernels on CT images will change the CT numbers and electron density of tissues, conse...
متن کاملOptimization of clinical target volume delineation using magnetic resonance spectroscopic imaging (MRSI) in 3D conformal radiotherapy of prostate cancer
Background: For the purpose of individual clinical target volume assessment in radiotherapy of prostate cancer, MRSI was used as a molecular imaging modality with MRI and CT images. Materials and Methods: The images of 20 prostate cancer patients were used in this study. The MR and MRSI images were registered with CT ones using non-rigid registration technique. The CT based planning (BP), CT/MR...
متن کامل