Predicted and measured oxygen concentrations in the circle system using low fresh gas flows with oxygen supplied by an oxygen concentrator.
نویسندگان
چکیده
An oxygen concentrator based on a zeolitic molecular sieve has been used to supply piped oxygen in a District General Hospital; such oxygen contains 5% argon. If concentrator oxygen is used in the closed circuit, argon accumulates. The present investigation defines the extent and time course of argon accumulation in partially and fully closed breathing systems. A theoretical model is presented which predicts the extent and rate of inert gas accumulation. The predictions have been tested in a volunteer under laboratory conditions and in five anaesthetized patients. It is recommended that concentrator oxygen may be used for low flow anaesthesia of indefinite duration, provided the fresh oxygen flow to the circuit is at least twice the oxygen consumption. If the circuit is totally closed, periodic opening of the circuit is necessary.
منابع مشابه
Inspired oxygen concentrations in semiclosed circle absorber circuits with low flows of nitrous oxide and oxygen.
Fresh gas flows of 1-5 l./min containing oxygen concentrations of 25-50% in nitrous oxide were led into a semiclosed circle absorber circuit. The resulting inspired oxygen concentrations were measured under conditions of spontaneous respiration. The lower the fresh gas flow, the greater was the discrepancy between fresh gas and inspired oxygen concentration. No association between minute volume...
متن کاملIsoflurane in a circle system with low gas flow.
Isoflurane was administered from a Goldman vaporizer to a circle breathing system with low fresh gas flows. The uptake and vapour concentration of isoflurane were compared in two groups of spontaneously breathing patients receiving oxygen alone (n = 19), or nitrous oxide and oxygen (n = 30). Anaesthesia was maintained successfully in all patients. The authors believe this to be a safe and econo...
متن کاملEffect of N2O on sevoflurane vaporizer settings during minimal- and low-flow anesthesia.
BACKGROUND Uptake of a second gas of a delivered gas mixture decreases the amount of carrier gas and potent inhaled anesthetic leaving the circle system through the pop-off valve. The authors hypothesized that the vaporizer settings required to maintain constant end-expired sevoflurane concentration (Etsevo) during minimal-flow anesthesia (MFA, fresh gas flow of 0.5 l/min) or low-flow anesthesi...
متن کاملDomiciliary oxygen concentrators in anaesthesia: preoxygenation techniques and inspired oxygen concentrations.
A conventional preoxygenation technique was compared in two volunteers with two techniques based on a drawover breathing system supplied with oxygen enriched air from a Devilbiss Mini VO2 domiciliary oxygen concentrator. Breathing from a 20-litre bag, filled previously at 2 litre min-1 from the concentrator was as effective as the conventional technique (PaO2 73.2 kPa). Supplying 4 litre min-1 ...
متن کاملGCMC Glauber dynamics study for structural transitions in YBCOx (0<x<1), HTc system
We have chosen an Ising ASYNNNI (ASYmmetric Next Nearest Neighbor Interaction) model under a grand canonical regime to investigate structural phase transition from a high symmetric tetragonal (Tet) to a low symmetric orthorhombic in YBa2Cu3O6+x , 0<x<1, HTc system. Ordering process for absorbed oxygens from an external gas bath into the basal plane of the layered system is studied by Monte C...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- British journal of anaesthesia
دوره 61 4 شماره
صفحات -
تاریخ انتشار 1988