We compared the results obtained with six commercial MIC test systems (Etest, MicroScan, Phoenix, Sensititre, Vitek Legacy, and Vitek 2 systems) and three reference methods (agar dilution, disk diffusion, and vancomycin [VA] agar screen [VScr]) with the results obtained by the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference method for the detection of VA-intermediate Staphylococcus aureus (VISA). A total of 129 S. aureus isolates (VA MICs by previous BMD tests, <or=1 microg/ml [n = 60 strains], 2 microg/ml [n = 24], 4 microg/ml [n = 36], or 8 microg/ml [n = 9]) were selected from the Centers for Disease Control and Prevention strain collection. The results of BMD with Difco Mueller-Hinton broth were used as the standard for data analysis. Essential agreement (percent +/-1 dilution) ranged from 98 to 100% for all methods except the method with the Vitek Legacy system, for which it was 90.6%. Of the six commercial MIC systems tested, the Sensititre, Vitek Legacy, and Vitek 2 systems tended to categorize VISA strains as susceptible (i.e., they undercalled resistance); the MicroScan and Phoenix systems and Etest tended to categorize susceptible strains as VISA; and the Vitek Legacy system tended to categorize VISA strains as resistant (i.e., it overcalled resistance). Disk diffusion categorized all VISA strains as susceptible. No susceptible strains (MICs <or= 2 microg/ml) grew on the VScr, but all strains for which the VA MICs were 8 microg/ml grew on the VScr. Only 12 (33.3%) strains for which the VA MICs were 4 microg/ml grew on VScr. The differentiation of isolates for which the VA MICs were 2 or 4 microg/ml was difficult for most systems and methods, including the reference methods.
منابع مشابه
Susceptibility To Vancomycin In Staphylococcus Aureus Isolated From Patients Of Four University-Affiliated Hospitals In Tehran
Background and Objective: Vancomycin is frequently the antibiotic of choice for the treatment of infections caused by methicillin-resistant Staphylococcus aureus (MRSA). For the last years, the incidence of vancomycin-intermediate S. aureus (VISA) and vancomycin-resistant S. aureus (VRSA) has been increased in various parts of the world. The present study ...
متن کاملVancomycin-intermediate and -resistant Staphylococcus aureus: what the infectious disease specialist needs to know.
Ever since the first strain of Staphylococcus aureus with reduced susceptibility to vancomycin and teicoplanin was reported from Japan, there has been a lot of confusion regarding the laboratory and clinical approach to patients with infections due to S. aureus with reduced susceptibility to vancomycin. To date, 6 clinical infections with vancomycin-intermediate S. aureus (VISA) have been repor...
متن کاملReduced sensitivity of staphylococcus aureus to vancomycin
Introduction: Staphylococcus aureus is one of the most common causes of infections in hospitals and vancomycin is used for the treatment of methicillin resistant S, aureus infections. Therefore, imergence of Staphylococcus aureus with reduced sensitivity to vancomycin is a serious problem. The aim of this study was to investigate the prevalence of Staphylococcus aureus with reduced suscep...
متن کاملDetermination of vancomycin and methicillin resistance in clinical isolates of Staphylococcus aureus in hospitals of Ilam city
Introduction: In this study, using the phenotypic and genotypic methods, oxacillin susceptibility in Staphylococcus aureus (S. aureus) strains isolated from patients at two government hospitals in Ilam, Iran was tested. Materials and methods: Out of 200 S. aureus isolates from different human clinical specimens consisting of blood (31%), wound (20%), urine (21%), catheters (7%), sputum (12...
متن کاملPerformance of Vitek 2 for antimicrobial susceptibility testing of Staphylococcus spp. and Enterococcus spp.
Vitek 2 (bioMérieux, Inc., Durham, NC) is a widely used commercial antimicrobial susceptibility testing system. We compared MIC results obtained by Vitek 2 to those obtained by the Clinical and Laboratory Standards Institute (CLSI) broth microdilution (BMD) reference method for 134 staphylococcal and 84 enterococcal clinical isolates. Nineteen agents were evaluated, including all those availabl...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of clinical microbiology
دوره 47 7 شماره
صفحات -
تاریخ انتشار 2009