Dosimetric characteristics of a new linear accelerator under gated operation
نویسندگان
چکیده
Respiratory gated radiotherapy may allow reduction of the treatment margins, thus sparing healthy tissue and/or allowing dose escalation to the tumor. However, current commissioning and quality assurance of linear accelerators do not include evaluation of gated delivery. The purpose of this study is to test gated photon delivery of a Siemens ONCOR Avant-Garde linear accelerator. Dosimetric characteristics for gated and nongated delivery of 6-MV and 15-MV photons were compared for the range of doses, dose rates, and for several gating regimes. Dose profiles were also compared using Kodak EDR2 and X-Omat V films for 6-MV and 15-MV photons for several dose rates and gating regimes. Results showed that deviation is less than or equal to 0.6% for all dose levels evaluated with the exception of the lowest dose delivered at 25 MU at an unrealistically high gating frequency of 0.5 Hz. At 400 MU, dose profile deviations along the central axes in in-plane and cross-plane directions within 80% of the field size are below 0.7%. No unequivocally detectable dose profile deviation was observed for 50 MU. Based on the comparison with widely accepted standards for conventional delivery, our results indicate that this LINAC is well suited for gated delivery of nondynamic fields.
منابع مشابه
Commissioning and early experience with a new generation low-energy linear accelerator in zahedan
Introduction: Since commissioning beam data are treated as a reference and ultimately used by treatment planning systems, it is vitally important that the collected data are of the highest quality to avoid dosimetric and patient treatment errors that may subsequently lead to a poor radiation outcome. Beam data commissioning should be performed with appropriate knowledge and pr...
متن کاملDelivery efficiency of an Elekta linac under gated operation
In this study, we have characterized the efficiency of an Elekta linac in the delivery of gated radiotherapy. We have explored techniques to reduce the beam-on delay and to improve the delivery efficiency, and have investigated the impact of frequent beam interruptions on the dosimetric accuracy of gated deliveries. A newly available gating interface was installed on an Elekta Synergy. Gating s...
متن کاملEvaluation of a combined respiratory‐gating system comprising the TrueBeam linear accelerator and a new real‐time tumor‐tracking radiotherapy system: a preliminary study
A combined system comprising the TrueBeam linear accelerator and a new real-time, tumor-tracking radiotherapy system, SyncTraX, was installed in our institution. The goals of this study were to assess the capability of SyncTraX in measuring the position of a fiducial marker using color fluoroscopic images, and to evaluate the dosimetric and geometric accuracy of respiratory-gated radiotherapy u...
متن کاملA study on needed or unneeded ReDosimetry and Recomissioning of a linac with Add on MLC by Comparing Dosimetric Data
Introduction: Today different uses of multileaf collimators(MLC) in different techniques of radiotherapy are common. Due to the use of this facility it is possible to have maximum radiation dose to target volumes and minimum radiation dose to critical tissues surronding treatment volumes. But there are numerouse Linear accelerators in IRAN without MLC. Because of high price of...
متن کاملEvaluation of dosimetric characteristics of a grid block fabricated for Mega-voltage grid therapy purposes
Background: In conventional radiation therapy, regarding normal tissue tolerance, the treatment of bulk tumors is one of the remaining challenges. Grid Radiation Therapy (GRT) is a technique to deliver high doses, approximately 15 – 20 Gy per fraction, to several small volumes located in a large radiation field. This can be performed using a grid block. The current work has concentrate...
متن کامل