Influence of the sliding velocity and the applied potential on the corrosion and wear behavior of HC CoCrMo biomedical alloy in simulated body fluids.
نویسندگان
چکیده
The corrosion and tribocorrosion behavior of an as-cast high carbon CoCrMo alloy immersed in phosphate buffered solution (PBS) and phosphate buffered solution with bovine serum albumin (PBS+BSA) have been analyzed by electrochemical techniques and surface microscopy. After the electrochemical characterization of the alloy in both solutions, the sample was studied tribo-electrochemically (by open circuit potential, OCP measurements, potentiodynamic curves and potentiostatic tests) in a ball-on-disk tribometer rotating in different sliding velocities. The influence of solution chemistry, sliding velocity and applied potential on the corrosion and tribocorrosion behavior of the CoCrMo alloy has been studied. Anodic current density increases with sliding velocity but wear rate does not change at an applied anodic potential; on the other hand, BSA modifies the wear debris behavior (by agglomerating the debris formed by mechanical removal of particles) thus increasing the mechanical wear volume. Under cathodic conditions, cathodic current density also increases during mechanical contact while the wear rate decreases with sliding velocity and BSA lubricates the contact thus reducing the total wear volume with respect to the non-containing BSA solution. The work shows how the electrode potential critically affects the corrosion and tribocorrosion rates by increasing the wear coefficients at applied anodic potentials due to severe wear accelerated corrosion.
منابع مشابه
Wear of CoCrMo alloys used in metal-on-metal hip joints A tribocorrosion appraisal
A good biocompatibility, excellent mechanical properties and high corrosion resistance characterize CoCrMo alloys. Therefore they are widely used for artificial joints in biomedical implants. However, the degradation of the implants during service life leads to the release into the body of toxic ions and wear particles. This continuous degradation is of concern for long-term stability of the im...
متن کاملEffect of Repeated Repair Metal Inert Gas Welding on Microstructural Properties, Corrosion Resistance, and Wear Behavior of 5083-H116 Aluminum Alloy
The effect of repeated repair welding, shielded with argon, on microstructural properties, corrosion resistance, and dry sliding wear behavior of aluminum alloy 5083/H116 were investigated. Samples were welded by metal inert gas welding method. 100% argon was used to protect fusion zone. Aluminum alloy 5356 was used as the filler metal. The samples for microstructure, corrosion, and wear tests ...
متن کاملEFFECT OF ANODIC OXIDATION ON THE CORROSION BEHAVIOR OF NICKEL-TITANIUM SHAPE MEMORY ALLOYS IN SIMULATED BODY FLUIDS (SBF)
The effect of anodic oxidation of a NiTi shape memory alloy in sulfuric acid electrolyte on its surface characteristics was studied. Surface roughness was measured by roughness tester. Surface morphology was studied using optical microscopy (OM) and scanning electron microscopy (SEM). Corrosion behavior was specified by recording Potentiodynamic polarization cur...
متن کاملWear-Corrosion Synergism Influences the Performance of High-Carbon and Low-Carbon CoCrMo
INTRODUCTION Recently, the popularity of Metal-on-Metal (MoM) hip bearings have diminished due to the concerns with generated wear debris and metal ion release that cause a negative response in the surrounding tissues. MoM bearings are typically manufactured from two types of metal alloys, lowcarbon (LC) and high-carbon (HC) CoCrMo. However, little is known about the in-vivo behavior of the two...
متن کاملEFFECTS OF Al-5TI-1B MASTER ALLOY AND HEAT TREATMENT ON THE MICROSTRUCTURE AND DRY SLIDING WEAR BEHAVIOR OF AN Al-12Zn-3Mg-2.5Cu ALLOY
In this study the effect of Al–5Ti–1B grain refiner on the structural characteristics and wear properties of Al–12Zn–3Mg–2.5Cu alloy was investigated. The optimum amount for Ti containing grain refiners was selected as 2 wt.%. T6 heat treatment, (i.e. heating at 460 °C for 1 h before water quenching to room temperature and then aging at 120 °C for 24 h) was applied for all specimens before wear...