Respiratory inductive plethysmography: a comparative study between isovolume maneuver calibration and qualitative diagnostic calibration in healthy volunteers assessed in different positions.
نویسندگان
چکیده
OBJECTIVE To compare two methods of respiratory inductive plethysmography (RIP) calibration in three different positions. METHODS We evaluated 28 healthy subjects (18 women and 10 men), with a mean age of 25.4 ± 3.9 years. For all of the subjects, isovolume maneuver calibration (ISOCAL) and qualitative diagnostic calibration (QDC) were used in the orthostatic, sitting, and supine positions. In order to evaluate the concordance between the two calibration methods, we used ANOVA and Bland-Altman plots. RESULTS The values of the constant of proportionality (K) were significantly different between ISOCAL and QDC in the three positions evaluated: 1.6 ± 0.5 vs. 2.0 ± 1.2, in the supine position, 2.5 ± 0.8 vs. 0.6 ± 0.3 in the sitting position, and 2.0 ± 0.8 vs. 0.6 ± 0.3 in the orthostatic position (p < 0.05 for all). CONCLUSIONS Our results suggest that QDC is an inaccurate method for the calibration of RIP. The K values obtained with ISOCAL reveal that RIP should be calibrated for each position evaluated.
منابع مشابه
Respiratory monitoring by inductive plethysmography in unrestrained subjects using position sensor-adjusted calibration.
BACKGROUND Portable respiratory inductive plethysmography (RIP) is promising for noninvasive monitoring of breathing patterns in unrestrained subjects. However, its use has been hampered by requiring recalibration after changes in body position. OBJECTIVES To facilitate RIP application in unrestrained subjects, we developed a technique for adjustment of RIP calibration using position sensor f...
متن کاملMathematical assessment of qualitative diagnostic calibration for respiratory inductive plethysmography.
We present a critical assessment of qualitative diagnostic calibration (QDC), which claims to provide a relative calibration of respiratory inductive plethysmography during natural breathing (Sackner MA, Watson H, Belsito AS, Feinerman D, Suarez M, Gonzalez G, Bizousky F, and Krieger B. J Appl Physiol 66: 410-420, 1989). QDC computes the calibration factor (K) by considering breaths of constant...
متن کاملRespiratory inductance plethysmography in healthy infants: a comparison of three calibration methods.
Respiratory inductance plethysmography (RIP) measures respiration from body surface movements. Various techniques have been proposed for calibration in order that RIP may be used quantitatively. These include calculation of the proportionality constant of ribcage to abdominal volume change (K). The aims of this study were to 1) establish whether a fixed value of K could be used for calibration,...
متن کاملRespiratory inductive plethysmography measurements on professional reed woodwind instrument players
Respiratory movements and lung volume variations during natural performance of wind instrument playing have been scarcely documented in the literature, but may provide a deeper insight into performance techniques, players’ physiological characteristics as well as into the physics of the instruments. Using Respiratory Inductive Plethysmography (RIP) respiratory movements of eight professional pl...
متن کاملEvaluation of respiratory inductive plethysmography in the measurement of breathing pattern and PEEP-induced changes in lung volume.
STUDY OBJECTIVE To assess the accuracy of the respiratory inductive plethysmography in the measurement of PEEP-induced changes in end-expiratory lung volume during mechanical ventilation and its accuracy and stability in the measurement of ventilation during controlled mechanical ventilation and spontaneous breathing. DESIGN An open comparison between two methods using a criterion standard. E...