Corneal flap thickness with the Moria M2 single-use head 90 microkeratome in 72 consecutive LASIK procedures

نویسندگان

  • Yunus Karabela
  • Orkun Muftuoglu
  • Faruk Kaya
چکیده

PURPOSE This study aimed to evaluate the accuracy and consistency of corneal flap thickness in laser-assisted in situ keratomileusis (LASIK) with the Moria M2 single-use head 90 microkeratome. METHODS The central corneal thickness of 72 (37 right, 35 left) eyes of 37 patients was measured by ultrasonic pachymetry preoperatively and intraoperatively after flap cut. The Moria M2 single-use head 90 microkeratome was used to create a superior hinged flap in all eyes. The right eyes were always operated on before the left eyes in each patient, using the same blade in all bilateral cases. All patients underwent LASIK for myopia and/or myopic astigmatism using VISX Star S4 platform. RESULTS The mean preoperative spherical equivalent refraction was -3.55±2.30 D (range: -0.625 to -11.00 D), preoperative central corneal thickness by ultrasonic pachymetry was 541±26.82 µm (490-600 µm) and steepest K was 44.08±1.49 D (40-46.75 D) in all eyes. The mean flap thickness was 136.97±20.07 µm (106-192 µm), 131.2±19.5 µm (91-192 µm), and 134.16±19.85 µm (91-192 µm) in the right, left, and both eyes, respectively. A positive significant relationship was found between flap thickness and preoperative ultrasonic pachymetry thickness. No significant relationship was found between flap thickness and the age, preoperative spherical equivalent, and preoperative steepest K. The difference between the first and second eyes was not significant. There were no major intraoperative and postoperative complications in all eyes. CONCLUSION The Moria M2 single-use head 90 microkeratome cut relatively thicker flaps than were intended. The flap thickness range was quite wide. This was a disadvantage for the accuracy and consistency of corneal flap thickness.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Corneal flap thickness with the Moria M2 single-use head 90 microkeratome.

PURPOSE To analyse the accuracy of corneal flap thickness created in laser-assisted in situ keratomileusis (LASIK) using the Moria Model 2 (M2) single-use head 90 microkeratome. METHODS The corneal thickness of 300 (266 myopic and 34 hyperopic) eyes of 150 patients was measured by ultrasonic pachymetry preoperationally and intraoperationally after flap cut. The Moria M2 single-use head 90, in...

متن کامل

The effect of flap thickness on corneal biomechanics after myopic laser in situ keratomileusis using the M-2 microkeratome

Purpose The purpose of this study was to determine the effect of flap thickness on corneal biomechanics after myopic laser in situ keratomileusis (LASIK). Methods This is a prospective controlled non-randomized, institutional study. Patients underwent either epi-LASIK with mitomycin (advanced surface ablation [ASA]), thin-flap LASIK (90 µm head), or thick-flap LASIK (130 µm head). In ASA, the...

متن کامل

Refinement of Intraocular Pressure Measurements made by Ocular Response Analyzer following Laser in Situ Keratomileusis using M2 90 Moria Microkeratome for Egyptian Myopic and Astigmatic Patients

INTRODUCTION The Goldmann applanation tonometer readings were noted to be markedly reduced after Lasik surgery using a thin flap technique which is widely used nowadays, to correct a wide range of myopia and astigmatism. The Ocular Response Analyzer (ORA) is considered one of the important tools for refinement of Intraocular Pressure (IOP) measurements. The aim of this study is to evaluate and ...

متن کامل

Comparison of corneal flap thickness using a FS200 femtosecond laser and a moria SBK microkeratome.

AIM To evaluate differences in flap thickness resulting from use of an Alcon Wavelight FS200 femtosecond laser and a MORIA SBK microkeratome when making a 110-µm-thick corneal flap and to identify the potential factors that affect corneal flap thickness. METHODS A prospective case study was performed on 120 eyes of 60 patients who were divided into two groups for LASIK, each group consisting ...

متن کامل

Temporal and Spatial Flap Variability in Laser In-Situ Keratomileusis by Optical Coherence Tomography

PURPOSE To study changes in flap thickness made with two different microkeratome heads across different corneal locations using anterior segment optical coherence tomography (OCT). METHODS In this prospective, non-randomized, consecutive case series, subjects who had their laser in-situ keratomileusis (LASIK) flaps made using 90 μm (MSU90) or 130 μm (MSU130) disposable M2 microkeratome heads ...