Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
نویسندگان
چکیده
PURPOSE To measure the interfraction setup variation of patient undergoing intensity-modulated radiation therapy (IMRT) of head and neck cancer. The data was used to define adequate treatment CTV-to-PTV margin. MATERIALS AND METHODS During March to September 2006, data was collected from 9 head and neck cancer patients treated with dynamic IMRT using 6 MV X-ray beam from Varian Clinac 23EX. Weekly portal images of setup fields which were anterior-posterior and lateral portal images were acquired for each patient with an amorphous silicon EPID, Varian aS500. These images were matched with the reference image from Varian Acuity simulator using the Varis vision software (Version 7.3.10). Six anatomical landmarks were selected for comparison. The displacement of portal image from the reference image was recorded in X (Left-Right, L-R), Y (Superior-Inferior, S-I) direction for anterior field and Z (Anterior-Posterior, A-P), Y (S-I) direction for lateral field. The systematic and random error for individual and population were calculated. Then the population-based margins were obtained. RESULTS A total of 135 images (27 simulation images and 108 portal images) and 405 match points was evaluated. The systematic error ranged from 0 to 7.5 mm and the random error ranged from 0.3 to 4.8 mm for all directions. The population-based margin ranged from 2.3 to 4.5 mm (L-R), 3.5 to 4.9 mm (S-I) for anterior field and 3.4 to 4.7 mm (A-P), 2.6 to 3.7 mm (S-I) for the lateral field. These margins were comparable to the margin that was prescribed at the King Chulalongkorn Memorial Hospital (5-10 mm) for head and neck cancer. CONCLUSION The population-based margin is less than 5 mm, thus the margin provides sufficient coverage for all of the patients.
منابع مشابه
Utilization of Electronic Portal Imaging Device (EPID) For Setup Verification and Determination of Setup Margin in Head and Neck Radiation Therapy
Introduction: Radiation therapy involves a multistep procedure; therefore, the error in patient set up is an inherent part of the treatment. Main purpose of this study was to determine the clinical target volume (CTV) to planning target volume (PTV) in head and neck cancer patients. Material and Methods: A total of 15 patients who had daily p...
متن کاملDose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...
متن کاملPretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)
Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...
متن کاملEffects of error on dose of target region and organs at risk in treating nasopharynx cancer with intensity modulated radiation therapy
OBJECTIVE To measure setup error of head and neck neoplasm in radiotherapy and discuss over effects of error on physical dose acting on target region and organs at risk of nasopharynx cancer (NPC) patients treated with intensity modulated radiation therapy (IMRT). METHODS A total of 152 patients who developed head and neck neoplasm and received IMRT were randomly selected. Through comparing d...
متن کاملComprehensive end-to-end test for intensity-modulated radiation therapy for nasopharyngeal carcinoma using an anthropomorphic phantom and EBT3 film
Background: In head and neck radiotherapy, immobilization devices can affect dose delivery. In this study, a comprehensive end-to-end test was developed to evaluate the accuracy of radiotherapy treatment. Materials and Methods: An Alderson Radiation Therapy (ART) anthropomorphic phantom with EBT3 film was used to mimic the actual patient treatment process. Ten patients treated for nasopharyngea...
متن کامل