Performance of the Agilent 1290 Infinity Quaternary Pump using Trifluoroacetic Acid (TFA) as the mobile phase modifier
نویسنده
چکیده
In liquid chromatography, baseline noise should be as low as possible to provide an optimum signal-to-noise ratio and to enable reliable integration of peaks at trace level. Baseline noise has different sources. One source is the detector, and another is the mixing performance of the pump, which is especially important if, for example, trifluoroacetic acid (TFA) is used as modifier in the mobile phase. This Technical Overview shows the mixing performance of the Agilent 1290 Quaternary Pump using TFA as the modifier in the mobile phase. Mixing performance depends strongly on the mixing volume. Typically, larger mixing volumes are better for higher baseline performance. Therefore, the influence of an additional mixer (V380 Jet Weaver) on the baseline noise of the 1290 Infinity Quaternary Pump is evaluated. The results obtained using the Agilent 1260 Infinity Quaternary Pump are compared with results obtained using a non-Agilent quaternary UHPLC system. Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 With mixer w/o mixer Pe ak -t ope ak n oi se (m A U ) Comparison of baseline noise 0% ACN 1% ACN 3% ACN 5% ACN 10% ACN 15% ACN 20% ACN 30% ACN 50% ACN 80% ACN 100% ACN
منابع مشابه
Retention time precision of Agilent 1290 Infinity Quaternary Pump Analysis of clopidol using dynamically and premixed gradients
The most important performance parameter for an ultrahigh performance liquid chromatography (UHPLC) pump is precision of retention times. Highest precision is required to identify compounds by retention times and subsequently accurately quantify the analyzed sample compounds. Gradient analysis can be done using either premixed mobile phases applying gradients from 0 to 100% B, as described in s...
متن کاملHigh-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System
The Agilent 1290 Infinity LC System with ultraviolet/visible (UV/VIS) Diode Array detection (DAD) is used to analyze triglycerides in soybean oil under non-aqueous reversed phase gradient conditions. The Agilent 1290 Infinity LC System was used for the chromatographic separation of the sample on 3.0 and 2.1 mm id C18 columns, of various lengths, with 1.8-μm packing materials prepared in 600 bar...
متن کاملPerformance Characteristics of the Agilent 1290 Infinity Evaporative Light Scattering Detector
The Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD) provides a wide detection range for a variety of compounds with excellent sensitivity, down to low-nanogram levels, better efficiency and greater reproducibility. It is perfect for the analysis of semivolatile and nonvolatile compounds and compounds without UV chromophores such as most amino acids and sugars. The 1290 Infini...
متن کاملFast screening of mobile and stationary phases with the Agilent 1290 Infinity LC and seamless method transfer to an Agilent 1200 Series LC using ISET
This Application Note describes how the Agilent 1290 Infinity LC System Method Development Solution was used to find optimum separation conditions for degradation products of a pharmaceutical preparation. Having evaluated the optimum combination, a transfer to conventional conditions was performed. Finally, the Agilent Intelligent System Emulation Technology (ISET) of the 1290 Infinity LC was u...
متن کاملEffect of Calcination Temperature and Li / TFA Molar Ratio on Morphology and Phase of Lithium Fluoride Nanoparticles Produced by Fluorolytic Sol-Gel Method
In this research, synthesis of lithium fluoride (LiF) nanoparticles by fluorolytic sol-gel method has been studied. Moreover, the effect of lithium ion to fluorine source molar ratio and calcination temperature on particle size and phase of LiF nanoparticles were investigated. Lithium acetate (C2H3LiO2), trifluoroacetic acid (TFA), ethylene glycol monobutyl ether and oleic acid were used as so...
متن کامل