Analysis of the optimum internal margin for respiratory‐gated radiotherapy using end‐exhalation phase assessments using a motion phantom
نویسندگان
چکیده
We aimed to optimize internal margin (IM) determination for respiratory-gated radiotherapy using end-expiratory phase assessments using a motion phantom. Four-dimensional computed tomography (4D CT) data were acquired using a GE LightSpeed RT CT scanner, a respiratory-gating system, and a motion phantom designed to move sinusoidally. To analyze the accuracy of 4D CT temporal resolution, a 25.4 mm diameter sphere was inserted into the motion phantom, and we measured the differences in sphere diameters between static and end-exhalation phase images. In addition, the IM obtained from the maximum intensity projection within the gating window (MIP(GW)) image was compared to theoretical value. Cranial-caudal motion displacement ranged from 5.0 to 30.0 mm, and the respiratory period ranged from 2.0 to 6.0 sec. Differences in sphere diameters between static and end-exhalation phase images ranged from 0.37 to 4.6 mm, with 5.0-mm and 30 mm target displacements, respectively. Differences between the IM obtained from the MIP(GW) and the theoretical values ranged from 1.12 to 6.23 mm with 5.0mm and 30 mm target displacements, respectively. These differences increased in proportion to the target velocity due to a motion artifact generated during tube rotation. In this study, the IMs obtained using the MIPGW image were overestimated in all cases. We therefore propose that the internal target volume (ITV) for respiratory-gated radiotherapy should be determined by adding the calculated value to the end-exhalation phase image. We also demonstrate a methodology for subtracting motion artifacts from the ITV using a motion phantom.
منابع مشابه
Investigation of lung normal tissue doses in lung tumors radiation therapy using both gated and conventional radiotherapy
Introduction: In radiation therapy of lung tumors, respiratory motion causes target moving, so a larger margin is needed to cover the clinical target volume (CTV). With the margin increasing, a larger volume of normal tissue will be exposed to high-dose. In this study, dosimetric parameters of normal lung tissue were compared between gated and conventional radiotherapy (RT), u...
متن کاملبررسی دز رسیده به بافت سالم ریه در پرتودرمانی تومورهای ریوی به دو روش پرتودرمانی گیتد و معمولی
In radiation therapy of lung tumors, respiratory motion causes target moving, so a larger margin is needed to cover the CTV. With the margin increasing, a larger volume of normal tissue will be exposed to high-dose. In this study, dosimetric parameters of normal lung tissue were compared between gated and conventional radiotherapy (RT), using the NCAT digital phantom. Different series of CT im...
متن کاملDose profile measurements during respiratory- gated lung stereotactic radiotherapy: A phantom study
During stereotactic body radiotherapy, high radiation dose (~60 Gy) is delivered to the tumour in small fractionation regime. In this study, the dosimetric characteristics were studied using radiochromic film during respiratory-gated and non-gated lung stereotactic body radiotherapy (SBRT). Specifically, the effect of respiratory cycle and amplitude, as well as gating window on the dosimetry we...
متن کاملTarget localization accuracy in a respiratory phantom using BrainLab ExacTrac and 4DCT imaging
This study evaluated the accuracy of measuring the motion of an internal target using four-dimensional computed tomography (4DCT) scanning and the BrainLAB ExacTrac X-ray imaging system. Displacements of a metal coil implanted in a commercial respiratory phantom were measured in each system and compared to the known motion. A commercial respiratory motion phantom containing a metal coil as a su...
متن کاملA Simulation Study on Patient Setup Errors in External Beam Radiotherapy Using an Anthropomorphic 4D Phantom
Introduction Patient set-up optimization is required in radiotherapy to fill the accuracy gap between personalized treatment planning and uncertainties in the irradiation set-up. In this study, we aimed to develop a new method based on neural network to estimate patient geometrical setup using 4-dimensional (4D) XCAT anthropomorphic phantom. Materials and Methods To access 4D modeling of motion...
متن کامل