Fluid preinjection for microwave ablation in an ex vivo bovine liver model assessed with volumetry in an open MRI system.

نویسندگان

  • Federico Collettini
  • Hendrik Rathke
  • Bernhard Schnackenburg
  • Andreas Thomas
  • Liane Albrecht
  • Britta Suttmeyer
  • Martin Jonczyk
  • Felix Guettler
  • Ulf Teichgräber
  • Thomas Kroncke
  • Bernd Hamm
  • Maximilian de Bucourt
چکیده

PURPOSE We aimed to detect possible differences in microwave ablation (MWA) volumes after different fluid preinjections using magnetic resonance imaging (MRI). MATERIALS AND METHODS MWA volumes were created in 50 cuboid ex vivo bovine liver specimens (five series: control [no injection], 10 mL water, 10 mL 0.9% NaCl, 10 mL 6% NaCl, and 10 mL 12% NaCl preinjections; n=10 for each series). The operating frequency (915 megahertz), ablation time (7 min), and energy supply (45 watts) were constant. Following MWA, two MR sequences were acquired, and MR volumetry was performed for each sequence. RESULTS For both sequences, fluid preinjection did not lead to significant differences in MWA ablation volumes compared to the respective control group (sequence 1: mean MWA volumes ranged from 7.0±1.2 mm [water] to 7.8±1.3 mm [12% NaCl] vs. 7.3±2.1 mm in the control group; sequence 2: mean MWA volumes ranged from 4.9±1.4 mm [12% NaCl] to 5.5±1.9 mm [0.9% NaCl] vs. 4.7±1.6 mm in the control group). The ablation volumes visualized with the two sequences differed significantly in general (P < 0.001) and between the respective groups (control, P ≤ 0.001; water, P < 0.001; 0.9% NaCl, P < 0.001; 6% NaCl, P ≤ 0.001; 12% NaCl, P < 0.001). The volumes determined with sequence 1 were closer to the expected ablation volume of 8 mL compared to those determined with sequence 2. CONCLUSION For the fluid qualities and concentrations assessed, there is no evidence that fluid preinjection results in larger coagulation volumes after MWA. Because ablation volumes determined by MRI vary with the sequence used, interventionalists should gain experience in how to interpret postinterventional imaging findings (with the MR scanner, sequences, and parameters used) to accurately estimate the outcome of the interventions they perform.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

PET and MRI-guided focused ultrasound surgery for hypoxic-tissue ablation combined with radiotherapy in solid tumors

Background: The rationale was to develop an ablation approach to destroy regions of tumor resistant to radiation and thus reduce the time required for whole tumor ablation, while improving overall tumor control after radiotherapy. Materials and Methods: The system is composed of a micro positron emission tomography (mPET), 7T magnetic resonance imaging (MRI), and a customized MRI-compatible foc...

متن کامل

Fluid-modulated bipolar radiofrequency ablation: an ex-vivo evaluation study.

BACKGROUND The effect of radiofrequency ablation (RFA) can be modulated by fluid injection. PURPOSE To evaluate the potential of different fluids to modulate the effect of bipolar RFA in an ex-vivo liver model. MATERIAL AND METHODS A bipolar RFA system with an internally cooled needle-shaped applicator (20-mm active tip) was used to induce coagulation lesions in ex-vivo porcine liver. Prior...

متن کامل

An Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation

Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...

متن کامل

Moving-Shot versus Fixed Electrode Techniques for Radiofrequency Ablation: Comparison in an Ex-Vivo Bovine Liver Tissue Model

OBJECTIVE To compare the ablation characteristics of the moving-shot technique (MST) and the fixed electrode technique (FET) for radiofrequency (RF) ablation in an ex-vivo bovine liver tissue model. MATERIALS AND METHODS We performed RF ablation using FET in 110 bovine liver blocks using 11 different ablation times ranging from 5 seconds to 5 minutes (10 blocks per each time duration). Ten bo...

متن کامل

Measurement of Intrahepatic Pressure during Microwave Ablation in an Ex Vivo Bovine Liver Model

BACKGROUND/AIMS We experimented with different ablation methods and two types of microwave antennas to determine whether microwave ablation (MWA) increases intrahepatic pressure and to identify an MWA protocol that avoids increasing intrahepatic pressure. METHODS MWA was performed using either a single-step standard ablation or a stepwise increment ablation paired with either a 16-gauge (G) 2...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Diagnostic and interventional radiology

دوره 19 5  شماره 

صفحات  -

تاریخ انتشار 2013