Impact of reorientation algorithms on quantitative myocardial SPECT perfusion imaging.
نویسندگان
چکیده
UNLABELLED In myocardial SPECT perfusion imaging, reorientation algorithms from transaxial image planes are used to generate short- and long-axis views of myocardial tracer uptake. We performed phantom experiments with 201Tl to delineate how image reorientation affects the results of quantitative image analysis. METHODS Thirty consecutive patient studies were analyzed to characterize the distribution of the angle of reorientation in a clinical setting. Short-axis SPECT images of a cardiac phantom with and without a 180 degrees cold-spot insert were reconstructed with three different backprojection filters (ramp, Metz and Butterworth) and reoriented through different angles ranging from 45 degrees to 89 degrees. Four interpolation algorithms were used to calculate from the transaxial images the pixel values of the reoriented images: (a) a simple interpolator that averages the pixel values of the eight neighboring pixels of the transaxial image; (b) a three-dimensional linear interpolator; (c) a hybrid interpolator that combines a two-dimensional linear in-plane with a one-dimensional cubic across-plane interpolation; and (d) a three-dimensional cubic convolution interpolator. Images were reoriented twice with opposite angles so that the original and the reoriented images could be directly compared. Circumferential profile analysis was applied to determine the root mean square error of corresponding profiles and the difference of the extent and the severity of perfusion defects. Single and multivariate analyses of variance (ANOVA) were used to compare the effects of the reorientation angle, the backprojection filter and the interpolation algorithm. RESULTS In the clinical studies, the angle between the transaxial and reoriented images was 75 degrees +/- 10 degrees (s.d.). In 48 phantom experiments, multivariate ANOVA demonstrated that the backprojection filter and the interpolation algorithm significantly affect the circumferential profiles and the extent and severity of a perfusion defect (p < 0.05). In contrast, the angle of reorientation was not a significant factor (p = ns). By univariate analysis, the three-dimensional cubic interpolator was associated with significantly (p < 0.05) less error than the simple and three-dimensional linear algorithms. Relative computation times (simple interpolator = 100%) were 119% for the three-dimensional linear, 136% for the hybrid and 243% for the three-dimensional cubic interpolator. CONCLUSION For quantitative analysis of myocardial SPECT perfusion images, a Metz filter for filtered backprojection in combination with a three-dimensional cubic convolution interpolation for image reorientation appears to offer improved accuracy.
منابع مشابه
Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms
Objective(s): A novel IQ-SPECTTM method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECTTM (IQ-mode),conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms. Materials and Me...
متن کاملComparison of transient ischemic dilation ratios in SPECT and SPECT-CT myocardial perfusion imaging in the low pre-test probability group
Introduction: The main purpose of this study was to compare transient ischemic dilation (TID) ratios in SPECT-low dose CT and SPECT Myocardial Perfusion Imaging (MPI) by application of different quantitative programs and quantify the possible shift in the upper normal limits of TID ratio in the SPECT-CT MPI. Methods: 149 Patients with low pre-test probability for coronary artery disease (CAD),...
متن کاملEvaluation of the potential impact of reconstruction method on dyssynchrony parameters derived by phase analysis of gated-SPECT MPI: Comparison of two quantitative software
Introduction: Gated SPECT myocardial perfusion scanning has new capabilities in addition to its main applications such as left ventricular dyssynchrony using phase analysis. Phase analysis has been investigated through various software including Emory Cardiac Toolbox (ECTb) and Quantitative Gated SPECT (QGS). The aim of this study is to evaluate the effect of reconstruction par...
متن کاملPerformance of myocardial perfusion imaging using multi-focus fan beam collimator with resolution recovery reconstruction in a comparison with conventional SPECT
Objective: IQSPECT is an advanced high-speed SPECT modality for performing myocardial perfusion imaging (MPI), which uses a multi-focus fan beam collimator with resolution recovery reconstruction. The aim of this study was to compare IQSPECT compared with conventional SPECT interms of performance based on standard clinical protocols. In addition, we examined the concordance between convention...
متن کاملFilter selection for 99mTc-Sestamibi myocardial perfusion SPECT imaging [Persian]
Myocardial perfusion SPECT imaging with Tc99m-sestamibi is the most accurate non-invasive means of detecting coronary artery disease and assessing the severity of perfusion abnormalities in the patients with coronary stenosis. Though simple and straight forward the results produced by the technique is very much affected by the details being used. One of the main problems with the test, wh...